Test function Voltage endurance test
Testing voltage 0 to 2.5 kV / 0 to 5.0 kV
AC, dual-range configuration
Voltage testing method Zero-toggle switch
Transformer capacity 500 VA (maximum 30
minutes)
Voltage adjustment Manually adjusted slidac
Waveform Mains waveform
Frequency Mains synchronous
Measurement items Voltage: 0 to 5 kV AC
(Average value rectified, effective value display), Current: 0.01 to 120
mA AC (Average value rectified, effective value display)
Judgment function Contents: UPPER-FAIL /
PASS / LOWER-FAIL (Digital setting window comparator method)
Timer section Setting range: 0.5 to 999
seconds
Interfaces EXT I/O, EXT SW, RS-232C
Display Fluorescent tube display (digital),
Analog meter
Monitor function Output voltage, detection
current
Power supply 120 V AC, 50/60 Hz (3158-01),
220 V AC, 50/60 Hz (3158-03), 230 V AC, 50/60 Hz (3158-04),240 V AC,
50/60 Hz (3158-05)
Dimensions, mass 320 mm(12.60 in)W × 155
mm(6.10 in)H × 263 mm(10.35 in)D
3158-01: 16 kg (564.4 oz), 3158-03/-04/-05:
18 kg (634.9 oz)
Accessories H.V. TEST LEAD 9615 (high
voltage side and return, 1 each), Power cord (1), spare fuse
|