D.C.
POWER SUPPLY ( < 100VA )
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
FEATURES: |
| |
||
| SPECIFICATIONS | ||
|
CONSTANT VOLTAGE OPERATION |
Regulation | Line regulation < 0.01% + 3mV |
| Load regulation < 0.01% + 3mV (rating current < 3A) | ||
| < 0.02% + 5mV (rating current > 3A) | ||
| Ripple & Noise | < 0.5mVrms 5Hz ~ 1MHz (rating current < 3A) | |
| < 1mVrms 5Hz ~ 1MHz (rating current > 3A) | ||
| Recovery Time | < 100µS (50% Load change, Minimum load 0.5A) | |
| Temp. Coefficient | < 300ppm/°C | |
| Output Range | 0 to rating voltage continuously adjustable | |
|
|
||
|
CONSTANT CURRENT OPERATION |
Regulation | Line
regulation < 0.2% + 3mA Load regulation < 0.2% + 3mA |
| Ripple Current | < 3mArms | |
| Output Range | 0
to rating amperes continuously adjustable (Hi / Lo range switchable) |
|
|
|
||
| METER | Analog | V-meter
and I-meter each one 2.5 class Dimension 50 x 50 mm |
| Digital | 3
1/2 digits 0.5" LED display 3 1/2 digits 0.39" LED display (for GPS-3030DD) Accuracy + (0.5% of rdg + 2 digits) |
|
|
|
||
| INSULATION | Chassis and Terminal | 20MΩ or above ( DC 500V ) |
| Chassis and AC Cord | 30MΩ or above ( DC 500V ) | |
|
|
||
| POWER SOURCE | AC 100V/120V/220V/240V+10%, 50/60Hz | |
|
|
||
| ACCESSORIES | Instruction
manual x 1, Power cord x 1 Test lead GTL-105 x 1 (<3A) or GTL-104 x 1 (<10A) |
|
|
|
||
| DIMENSIONS | 128(W) x 145(H) x 285(D) mm | |
SELECTION GUIDE |
| Model | Output Volts (V) | Output Amps (A) | ||
|
Digital |
GPS-1830D | 0 ~ 18 | 0 ~ 3 | |
| GPS-1850D | 0 ~ 18 | 0 ~ 5 | ||
| GPS-3030D | 0 ~ 30 | 0 ~ 3 | ||
| GPS-3030DD | 0 ~ 30 | 0 ~ 3 |
|