Diamond AccuPro Multi-function Universal Calibrator
More Calibrators
* 13 thermocouples
* 5 RTD's
* Frequency
* Trip detection
* mV
* Voltage
* Resistance
* Loop simulation
* Simultaneous input and output
* 21 step setpoints
* Auto/Manual setpoint cycling
* Ramping
* 24 VDC loop power
* Sealed keypad
* Full alpha-numeric keypad
* Carrying case, full lead set
* 3 year warranty
Optional Features
Pressure capability (CL-9001/CL-9002):
Enables the calibrators to accept pressure signals from the PMM-9700 and PM-9850 series
modules
Documenting/Delegated calibrators (CL-9002/CL-9004):
Provide the ability to store up to 50 delegated or data logging calibrations which can be
sent to
a terminal or serial printer, or interfaced to a computer aided calibration software
package.
CL-9001 Diamond Universal multi-function, pressure compatible calibrator
CL-9002 Diamond Plus Universal multi-function, pressure compatible and documenting
calibrator
CL-9003 Diamond Universal multi-function calibrator
CL-9004 Diamond Universal multi-function, documenting calibrator
Accessories
1079-009 120 VAC power adaptor
1079-010 240 VAC power adaptor
MC-1000
$$1295
High Accuracy Multi-Function Calibrator
COMMON
FEATURES
Excellent
accuracy ±0.4°C T/C ±0.3°C RTD and 0.015% electrical
Choose either
read or source modes
Operates with
external pressure modules
10 T/C types; 8
RTD types plus YSI 400 series thermistor (medical applications)
Full function 4
to 20 mA read, source, simulate and read/power modes with 0.015%
reading accuracy
Source 0 to 20
volts or read 0 to 20 volts with 0.015% of reading
Measure
0 to 250 VDC or VAC
Store
up to nine setpoints for each
RTD
function works with all pulsed XMTR’s
Direct
keyboard entry allows for quick output changes
RS232
Interface
A full multi-function
calibrator at a very competitive price with excellent accuracy best
describes the MC-1000. The MC-1000 combines the accuracy and functions
of our next generation PTC-8000 temperature calibrator with V/mA and
frequency functions added. The per-formance level is twice that of
similarly priced products and will calibrate virtually any process
device.
The MC-1000 also combines several DMM functions like voltage measurement
to 250 V AC or DC and a continuity test function. This combination of a
high accuracy calibrator along with common DMM functions makes the
MC-1000 the only instrument you may ever need to maintain and calibrate
your instrumentation.
Each unit includes: Test leads, carrying case, 4 AA alkaline batteries,
NIST Certificate, and instruction manual.
General
Specifications:
Pressure
Ranges:
Various
ranges in 29 different pressure modules – see Pressure Module
Ranges: Specifications
T/C
Ranges:
J,K,T,E,R,S,N,B,L,U
including -10 to70 mV
RTD
Ranges:
Pt
385 (100, 200, 500, 1000 ohms)
Pt 392, JIS, Ni120, CU10, YSI400
Ohms
Ranges:
0
to 400.00 and 400.0 to 4000.0
mA
Range:
0
to 24.000 mA
Voltage:
(Low)
Read 0-20.000 VDC Source 0-10.000 VDC
(High) Read only 0-250 VDC or VAC
Frequency:
0-1000.0
CPM
0-1000.0 Hz
0-10.00 KHz
Continuity:
0
to 400.0 ohms, audible tone when under 10 ohms
Accuracy:
T/C
Type J: ±.3°C ±10 m V
(±0.4°C total error)
RTD PT100: ±.1°C ±.075
(±0.3°C total error)
Ohms (400): ±0.1
Ohms (4000): ±1.0
mA
Compliance:
Up
to 1000
load
Voltage
Drive Capability:
1
mA
RTD
IEX-range:
0.01
mA to 3 mA
RTD
Frequency Response:
10
mS – works with all pulsed XMTR’s
Oper.
Temp:
-10°C
to 50°C
Storage
Temp:
-20°C
to 70°C
Loop
Power:
4
AA alkaline batteries, 25 hours typical operating time
Size:
7.5"
x 4" x 1.5"
More Calibrators