CMM15 Process Calibrator MultiMeter
Precision DCmA, Voltage, Frequency Calibrator combined with True RMS MultiMeter - 15
Functions
Features:
- Simultaneous source/simulate and measure on large dual display
- Precision 0.03% accurate voltage and current calibration source
- Constant current output for loads up to 500 ohms (Optional 1000 ohms adaptor)
- Bipolar current (±25mA) and voltage (±1.5V, ±15V) outputs
- Simultaneously source excitation voltage and measure mA output of 2-wire transmitters
- Wide range Frequency (square wave) source with adjustable amplitude, pulse width, and
duty cycle
- Programmable for 16 continuous or timed step outputs, or 999 step ramp
- Programmable 999 step ramp
- Optional Windows 95/3.1 software generates basic Calibration Report for unit under test
and enables graphing and datalogging
- Measure DCmA, ACmA, DCV, ACV, AC+DC, Temperature, Frequency, Duty Cycle, Pulse Width,
Resistance, Continuity, and Diode
- 1uV or 1uA resolution and 0.03% basic DCV accuracy (400mV-40V)
- 1ms Peak Hold for glitch capture, plus Data Hold, Auto Hold, relative and dynamic
recording
- True RMS measurement for non-linear voltage and current loads
- Built-in isolated RS-232 interface for data acquisition and documentation with optional
Windows data acquisition software
- Large 4,000/40,000 count backlit LCD displays primary, secondary, source/measure ranges
and functions selected
- Separate input/output banana jacks with fused protection
- Auto power off with disable function
- Complete with input/output test leads, Type K temperature probe, 9V battery, external
battery pack, protective holster and pouch case
|

|
| Source(Output) |
Ranges |
Max Resolution |
Accuracy±(%output+digits) |
| DC Voltage |
+/-1.5V, +/-15V |
100uV, 1mA |
(0.03%+0.3mV) |
| DC Current |
+/-25mA |
1uA |
(0.03%+5uA) |
| Square Wave(28 steps) |
0.5 to 4800Hz |
0.01Hz |
(0.005%+0.01Hz) |
| Measure(Input) |
Ranges |
Max Resolution |
Accuracy±(%rdg+digits) |
| DC Voltage |
40mV |
1uV |
(0.08%+5d) |
| DC Voltage |
400mV,4V,40V,300V |
10uV |
(0.03%+3d)* |
| AC Voltage(45-20kHz) |
40,400mV,4V,40V,300V |
1uV |
(0.7%+5d)@50/60Hz |
| DC Current |
40mA,400mA |
1uA |
(0.2%+3d) |
| AC Current(45Hz-2kHz) |
40mA,400mA |
1uA |
(1%+5d) |
| Resistance |
400,4k,40k,400k,4M ohms |
0.01 ohms |
(0.2%+3d) |
| Resistance |
40M ohms |
1k ohms |
(1%+5d) |
| Temperature(Type K) |
-40°F to 2192°F(-40°C to 1200°C) |
1/0.1°F(°C) |
(0.3%+6°F)/(0.3%+3°C) |
| Frequency |
100Hz,1,10,100,200kHz |
0.001Hz |
(0.2%+1d) |
| Duty Cycle |
0.1-99.9% |
|
(0.3% per kHz+0.3%)FS) |
| Pulse Width |
0.2-1999.9mS |
|
(0.2%+3d) |
| Dimensions |
1.45x3.54x7.56(37x90x192mm) |
Weight |
32oz.(940g) |
| *DCV accuracy (0.06%+3d)>40V |
and for negative readings |
|
|
|
CMM-15 |
CMM-15 Process Calibrator MultiMeter |
$499.00 |
|
149116 |
Optional 117VAC Adaptor 12VDC Source |
$15.00 |
|
423235 |
RS-232 Cable and Software diskette |
$69.00 |
|
|
|
|
|
CMM-13 Process Calibrator MultiMeter
Precision DC, mA, Voltage, Calibrators combined with True RMS MultiMeter
Features:
- True RMS measurement for non-linear voltage and current loads
- Built-in isolated RS-232 interface for data acquisition and documentation with optional
Windows data acquisition software
- Large 4,000/40,000 count backlit LCD displays primary, secondary, source/measure ranges
and functions selected
- Separate input/output banana jacks with fused protection
- Auto power off with disable function
- Calibrator Features:
- Simultaneous source/simulate and measure on large dual display
- Precision 0.03% accurate voltage and current calibration source
- Constant current output for loads up to 500* (Optional 1000* adaptor)
- Bipolar current (±25mA) and voltage (±1.5V, ±15V) outputs
- Simultaneously source excitation voltage and measure mA output of 2-wire transmitters
- Programmable for 16 continuous or timed step outputs, or 999 step ramp
- Optional Windows 95/3.1 software generates basic Calibration Report for unit under test
and enables graphing and datalogging
- DMM Features:
- Measure DCmA, ACmA, DCV, ACV, AC+DC, Temperature, Frequency, Duty Cycle, Pulse Width,
Resistance, Continuity, and Diode
- ACV and DCV to 600V
- 1µV or 1µA resolution and 0.03% basic DCV accuracy (400mV-40V)
- 1ms Peak Hold for glitch capture (CMM-15), plus Data Hold, Auto Hold. relative and
dynamic recording
|

|
|
CMM-13 |
CMM-13 Process Calibrator MultiMeter |
$499.00 |
|
423235 |
RS-232 cable and Windows software diskette |
$69.00 |
|
149116 |
Optional 117VAC Adaptor 12vdc Source |
$15.00 |
|
149230 |
Optional 220VAC Adaptor 12vdc Source |
$15.00 |
Digital Multimeters
|