The AirGard 200 continuously monitors fume hood face
velocity by measuring the air velocity of clean air from the room passing through the air
inlet on the front of the monitor. This simple, go/no-go alarm is a low cost option for
your monitoring requirements. It is an easy to install flush mount monitor that only
requires a single point calibration in the field. The device provides both visual and
audible alarms to alert the user of abnormal airflow conditions. A green indicator on the
front of the monitor indicates normal flow conditions. When flow conditions lower than the
setpoint are encountered, a red indicator is activated along with an audible horn. This
information can be sent to a remote location by means of a dry contact relay output.
SPECIFICATIONS:
Specifications Alarm Range 70 to 250 fpm (.35 to 1.25 m/s). Accuracy ±10% of set point or 10 fpm, whichever is greater. Alarm Delay 5 seconds. Audible Alarm Indication 85dB @ 4 inches (10 cm). Visual Alarm Indication Red LED. Relay Output 1.2A @ 12V DC Horn Silence Temporary & permanent. Input Selectable for either horn disable or alarm input. Calibration Alarm set point potentiometer adjustment. Instrument Dimensions
Front Faceplate 4.75L x 3.25W x .5D inch (12.07L x 8.25W x 1.27D cm).
Rear Enclosure 4.5L x 2.5W x 1.38D inch (11.43L x 6.35W x 3.50D cm). Mounting Flush. (Cut out required on hood surface). Operating Conditions 55 to 86°F (13 to 30°C), 5% to 95% RH, non-condensing. Storage Temperature -40 to 150°F (-40 to 65°C), 5% to 95% RH, non-condensing. Power Requirement (Wall plug in power supply supplied for domestic units).
Nominal Input Voltage 12.0 ± 1.0VDC or 9.0 ± 0.5VAC.
Nominal Input Current 0.12ADC @ 12VDC input.
Model
Description
Part Number
200
Flush mount. Red light/green light
indication. Low flow alarm.
Includes mounting hardware, tubing, sidewall adapter,
wall plug-in power supply and owner's manual.
670-200-120
List Price: $335
Alnor
AirGard® 335 Fume Hood Monitor
Price - $495
The AirGard 335 continuously monitors fume hood face
velocity by measuring the air velocity of clean air from the room passing through the air
inlet on the front of the monitor. The monitor then reports this information through
readily understandable means to the user. An electronic bargraph moves back and forth
through universally recognized red (danger), yellow (caution) and green (safe) colored
zones as the velocity changes. Corresponding large colored lights also illuminate. A
continuous audible alarm alerts of a dangerous situation when the face velocity is in an
alarm condition. This information can be sent to other personnel in a remote location by
means of a dry contact relay output. A digital display can be enabled to view velocity
quantitatively. The monitor mounts flush to the fume hood surface and fits into a single
electrical receptacle box knock-out.
SPECIFICATIONS: Digital Display 3 digit, 7-segment liquid crystal display with status
indicators and icons. Electronic bargraph moves back and forth through colored zones as
the velocity changes. Display Units If digits are enabled, velocity is represented in feet per
minute (fpm) or meters per second (m/s). User selectable through menu. Display Range 50 to 250 fpm (.25 to 1.27 m/s). Over range measurements
shown as Hi. Under range measurements shown as Lo. Resolution 1 fpm or .01 m/s. Accuracy ±10% or 10 fpm, whichever is greater. Alarm Range 50 to 250 fpm (.25 to 1.27 m/s). Alarm Delays User configurable through menu from 0 to 255 seconds. Audible Alarm Indication 85dB @ 4 inches (10 cm). Horn Silence Temporary & permanent. Night Set back Monitor can be disabled remotely. Visual Alarm Indication Jumbo red LED for low flow and high flow alarm
zone. Visual Warning Indication Jumbo yellow LED for low flow and high flow
warning zones. Warning offset is configurable through menu. Visual Normal Indication Jumbo green LED for normal flow zone. High Flow Alarm Disable Audible, visual and relay disable. Alarm Relay Output Nominal switching capacity 1A @ 30V DC, .5A @ 125V AC,
form A relay. Emergency Purge Button on front of monitor activates a relay that can be
used to command the fume hood damper to full open until pressed again. Nominal switching
capacity 1A @ 30V DC, .5A @ 125V AC, form A relay. External Alarm Input Intermittent audible and LCD status indicator
indicate external alarm. Sash Override Alarm Input Audible and LCD status indicator indicate the
sash position switch has been tripped. Calibration Two point field calibration required. Instrument Dimensions
Front Faceplate 5.25L x 3.0W x 5/8D inch (13.3L x 7.6W x 1.6D cm).
Rear Enclosure 3.0L x 2.0W x .5D inch (7.6L x 5.0W x 1.3D cm). Mounting Flush. (3.0L x 2.0W inch cut out required on hood surface).
Monitor fits into a single electrical receptacle box knock-out. Operating Conditions 55 to 86°F (13 to 30°C), 5% to 95% RH,
non-condensing. Storage Temperature -40 to 150°F (-40 to 65°C), 5% to 95% RH,
non-condensing. Power Requirement 9 to 30V AC/DC. (Wall plug in power supply supplied for
domestic units).
The calibration and configurable parameters are
stored to nonvolatile memory of the instrument and are not lost when the monitor loses
power.
Model
Description
Part Number
335
Flush mount. Colored analog scale and digital
flow indication.
Includes mounting hardware, tubing, sidewall adapter,
wall plug-in power supply and owner's manual.
670-335-000
List Price: $495
Alnor
AirGard® 385 Fume Hood Monitor
Price - $495
The AirGard 385 continuously monitors fume hood face
velocity by measuring the air velocity of clean air from the room passing through the air
inlet on the front of the monitor. The unit then reports this information visually through
a digital display and colored lights, as well as with a continuous audible alarm when the
face velocity is in an alarm condition. This information can be sent to a remote location
by means of a dry contact relay output and 0-4 volt analog output. The monitor mounts
flush to the fume hood surface and fits into a single electrical receptacle box knock-out.
SPECIFICATIONS:
Digital Display 3digit, 7-segment liquid crystal display. Display Units Feet per minute (fpm) or meters per second (m/s). User
selectable through menu. Display Range 70 to 250 fpm (.35 to 1.25 m/s). Over range measurements
shown as Hi. Under range measurements shown as Lo. Resolution 1 fpm or .01 m/s. Accuracy ±10% or 10 fpm, whichever is greater. Alarm Range 70 to 250 fpm (.35 to 1.25 m/s). Alarm Delays User configurable through menu from 0 to 255 seconds. Audible Alarm Indication 85dB @ 4 inches (10 cm). Visual Alarm Indication Red LED. Visual Warning Indications Yellow LEDs for low flow and high flow
warnings zones. Warning offsets are user configurable through menu. Relay Output Nominal switching capacity 1A @ 30V DC, .5A @ 125V AC, form
C relay. High Flow Alarm Disable Audible, visual and relay disable. Horn Silence Temporary & permanent. Night Set back Unit can be disabled remotely. User Alarm Input Intermittent audible and LCD status indicator indicate
external alarm. Analog Output 0 to 4V DC. Calibration Two point field calibration required. Instrument Dimensions Front Faceplate 5.25L x 3.0W x .5D inch (13.3L x 7.6W x 1.3D cm).
Rear Enclosure 3.0L x 2.0W x .5D inch (7.6L x 5.0W x 1.3D cm). Mounting Flush. (3.0L x 2.0W inch cut out required on hood surface).
Monitor fits into a single electrical receptacle box knock-out. Operating Conditions 55 to 86°F (13 to 30°C), 5% to 95% RH,
non-condensing. Storage Temperature -40 to 150°F (-40 to 65°C), 5% to 95% RH,
non-condensing. Power Requirement 9 to 24V AC/DC. (Wall plug in power supply supplied for
domestic units).
The calibration and configurable parameters are
stored to nonvolatile memory of the instrument and are not lost when the unit loses power.
Model
Description
Part Number
385
Flush mount. Digital flow indication (high and low alarm).
Includes mounting hardware, tubing, sidewall adapter,
wall plug-in power supply and owner's manual.
670-385-000
List Price: $495
Alnor
AirGard® 405 Fume Hood Monitor
Price - $335
The AirGard 405 continuously monitors fume hood face
velocity by measuring the air velocity of clean air from the room passing through the air
inlet on the front of the monitor. This simple, go/no-go alarm is a low cost option for
your monitoring requirements. It is an easy to install surface mount monitor that only
requires a single point calibration in the field. The device provides both visual and
audible alarms to alert the user of abnormal airflow conditions. A green indicator on the
front of the monitor indicates normal flow conditions. When flow conditions lower than the
setpoint are encountered, a red indicator is activated along with an audible horn. This
information can be sent to a remote location by means of a dry contact relay output.
SPECIFICATIONS: Alarm Range 70 to 250 fpm (.35 to 1.25 m/s). Accuracy ±10% of set point or 10 fpm, whichever is greater. Alarm Delay 5 seconds. Audible Alarm Indication 85dB @ 4 inches (10 cm). Visual Alarm Indication Red LED. Relay Output 1.2A @ 12V DC Horn Silence Temporary & permanent. Input Selectable for either horn disable or alarm input. Calibration Alarm set point potentiometer adjustment. Instrument Dimensions
Front Faceplate 6.0L x 3.13W x 1.75D inch (15.24L x 7.95W x 4.45D cm). Mounting Surface. (Monitor is mounted to metal backplate installed on
surface). Operating Conditions 55 to 86°F (13 to 30°C), 5% to 95% RH,
non-condensing. Storage Temperature -40 to 150°F (-40 to 65°C), 5% to 95% RH,
non-condensing. Power Requirement (Wall plug in power supply supplied for domestic
units).
Nominal Input Voltage 12.0 ± 1.0VDC or 9.0 ± 0.5VAC.
Nominal Input Current 0.12ADC @ 12VDC input.
Model
Description
Part Number
405
Surface mount. Red light/green light indication.
Low flow alarm.
Includes mounting hardware, tubing, sidewall adapter,
wall plug-in power supply and owner's manual.
670-180-010
List Price: $335
Alnor
AirGard® 480 Fume Hood Monitor
Price - $625
Call 800-783-6664 to order.
The AirGard 480 continuously monitors fume hood face
velocity by measuring the air velocity of clean air from the room passing through the air
inlet on the front of the monitor. The face velocity is displayed numerically, but the
monitor also features an LED column of lights to show air flow trends. The instrument is
flush mounted to the surface of the fume hood. It provides visual, audible, and remote
alarms for abnormal air flow conditions. This information can be sent to a remote location
by means of a dry contact relay output and 0-5 volt analog output. It can also print and
store air flow measurements at periodic intervals through the communications port when
connnected to a personal computer.
SPECIFICATIONS:
Model
Description
Part Number
480
Surface mount. Digital flow indication (high and low
alarm).
Includes mounting hardware, tubing, sidewall adapter,