Model
MC6 |
Multi-Cal Benchtop
Calibrator
Interchangeable Ranges, ± 0.05% Accuracy, Datalogging
Capability, NIST Traceable |
 |
|
Model MC6 Multi-Cal Benchtop Calibrator performs a wide
variety of simple and complex pressure based measurement, test, and
calibration operations. Modular sensor design allows user to select
pressure measurement range for application flexibility. Calibrator can
accommodate up to two interchangeable pressure modules (sold separately
below) in any combination of range or accuracy.
Simultaneously display two separate measurements on the two line,
alphanumeric display. Readings can be displayed in a choice of twelve
preprogrammed engineering units or any single user-defined unit.
Calibrator features min/max recall, hi/lo alarm, percentage of full
scale pressure readings, mA/voltage measurement, leak rate and pressure
decay measurement, switch testing capabilities, including trip point and
dead band, and velocity/volume flow rates.
Quickly document calibration procedures using the data logging
feature which stores up to 384 sets of pressure and time/date stamped
measurements. The automatic % error calculation simplifies transmitter
calibration. Easily upload stored data to an IBM compatible computer via
the RS232 two-way communications port. Calibrators and modules include
certification to NIST traceability--ideal as a secondary standard for
calibrating pressure equipment.
Multi-Cal Pressure Calibrator includes utility software, test leads,
protection module, adjustment handle, AC adapter, and instruction
manual.
Multi-Cal Pressure Modules are interchangeable and available
in a wide selection of pressure ranges and accuracies. Benchtop
calibrator accepts up to two pressure modules. Modules include NIST
calibration certification.
POPULAR
MODELS
Model
Number |
Description |
Price |
| MC6 |
BenchtopCalibrator |
$1625.00
|
POPULAR
MODELS
Model
Number |
Range |
Price |
| MC1000 |
0.25" H2O Differential Pressure, ±0.07 |
$1200.00 |
| MC1001 |
0.50" H2O Differential Pressure, ±0.07 |
$1200.00 |
| MC1004 |
5.00 H2O Differential Pressure, ±0.06 |
$980.00 |
| MC1006 |
25" H2O Differential Pressure, ±0.06 |
$980.00 |
| MC2010 |
5.0 psig Gauge Pressure, ±0.05 |
$780.00 |
| MC2012 |
15.0 psig Gauge Pressure, ±0.05 |
$780.00 |
| MC2016 |
100.0 psig Gauge Pressure, ±0.05 |
$780.00 |
OPTIONS- Consult us at 800-873-6664 for other pressure ranges
and accuracies. FM approved models are also available.
|
SPECIFICATIONS
Accuracy: Differential Pressure Modules: ± 0.06% FS,
Gauge Pressure Modules: ± 0.05% FS,
Voltage Input: ± 0.025% FS @ 0/10 VDC, ± 0.10% FS @ 0/30 VDC,
Current Input: ± 0.03% FS @ 0/20 mA, ± 0.05% FS @ 0/50 mA.
Repeatability: Ranges ≤ 0/2 psi: ± 0.05% of span,
Ranges ≥ 0/5 psi: ± 0.02% of span.
Sensitivity: ± 0.002% of span with dampening 1 part in 50,000
(max).
Overpressure: Differential Pressure Modules: 50 psi positive
direction,
15 psi negative direction, Gauge Pressure Modules: 2 x range (0/5 psi to
0/1000 psi).
Media: Clean, dry, nonconductive, noncorrosive gases.
Output: RS232 serial interface, 9-pin.
Baud rate: 300, 1200, 2400, or 9600, selectable.
Alarm Output: SPST from C 110 VDC, 120 VDC (max), 1 A (max), 30
W,
62.5 VA (resistive).
Dampening: (Measurement averaging) programmable from 0 to 16
consecutive readings.
Temperature Compensation: 20 to 120°F (-7 to 49°C).
Temperature Error: Maximum of ± 0.004% of span per °F over
compensated range
for zero and span.
Temperature Effect Electrical Measurement: ± 0.001% of span per
°F
over compensated range.
Ambient Operating Temperature: 32 to 120°F (0 to 49°C).
Storage Temperature: -4 to 158°F (-20 to 70°C).
Power Supply: 110 VAC/60 Hz adapter, 500 mA.
Engineering Units: inH2O, psi, inHg, kPa, mbar, bar,
cmH2O, mmHg, ftSW, MPa, mmH2O, kgcm2
and user-defined.
Display: Alphanumeric LCD, 0.37" (9.5 mm) height per line, 2
lines, 16 characters/line.
Display Update: 100 msec.
Computer Requirements: IBM compatible 386 or above and
Windows™ 3.1 or later with 2mB RAM and 2mB hard drive disk space, one
serial port.
Process Connection: 1/8" Female NPT.
Electrical Connections: Standard banana jacks.
Housing: ABS plastic.
Weight: Calibrator: 3.0 lb (1.4 kg), Pressure Module: 0.5 lb (0.3
kg).
APPLICATIONS:
Pressure measurement for clean room control, filter performance
monitoring, HVAC testing and setup, draft measurement, airflow
measurement control, differential pressure measurement in laminar
flowhoods, paint booths, industrial ovens and fume hoods. Use as a
secondary standard for calibrating pressure equipment.
|
 |
Series
MC |
Multi-Cal Pressure Calibrator
Interchangeable Ranges, ± 0.05% Accuracy, Datalogging
Capability, NIST Traceable |
 |
|
Series MC Portable Multi-Cal Pressure Calibrator performs a
wide variety of simple and complex pressure based measurement, test, and
calibration operations. Modular sensor design allows user to select
pressure measurement range for application flexibility. Calibrator can
accommodate up to two interchangeable pressure modules (sold separately
below) in any combination of range or accuracy.
Simultaneously display two separate measurements on the two line,
alphanumeric display. Readings can be displayed in a choice of seven
preprogrammed engineering units or any single user-defined unit.
Calibrator features min/max recall, hi/lo alarm, percentage of full
scale pressure readings, mA/voltage measurement, leak rate and pressure
decay measurement, switch testing capabilities, including trip point and
dead band, and velocity/volume flow rates.
Quickly document calibration procedures using the data logging
feature which stores up to 384 sets of pressure and time/date stamped
measurements. Easily upload stored data to an IBM compatible computer
via the RS232 port. Calibrators and modules include certification to
NIST traceability - ideal as a secondary standard for calibrating
pressure equipment.
Multi-Cal Pressure Calibrator includes utility software, test leads,
protection module, adjustable had strap, AC adapter, two 9V batteries,
instruction manual, and hard carrying case.
Multi-Cal Pressure Modules are interchangeable and available
in a wide selection of pressure ranges and accuracies. Handheld
calibrator accepts up to two pressure modules. Modules include NIST
calibration certification.
POPULAR
MODELS
Model
Number |
Description |
Price |
| MC2K |
Handheld Calibrator |
$1540.00 |
| CP40 |
Serial Printer |
$399.00 |
CP40 OPTIONS-
Consult factory for other pressure ranges and accuracies. FM approved
models are also available.
|
SPECIFICATIONS
Accuracy: Differential Pressure Modules: ±0.06% FS, Gauge
Pressure Modules: ±0.05% FS, Voltage Input: ±0.025% FS @ 1/10 VDC, ±0.10%
FS @ 0/30 VDC, Current Input: ±0.03% FS @ 1/20 mA, ±0.05% FS @ 0/50 mA.
Repeatability: Ranges ≤0/2 psi: ±0.05% of span, Ranges
≥0/5 psid: ±0.02% of span.
Sensitivity: ±0.002%of span with dampening 1 part in 50,000
(max.).
Overpressure: Differential Pressure Modules: 50 psi positive
direction, 15 psi negative direction, Gauge Pressure Modules: 2x range
(0/5 psi to 0/1000 psi.
Media: Clean, dry, nonconductive, noncorrosive gases.
Output: RS232 serial interface, 9-pin.
Baud Rate: 300, 1200, 2400, or 9600, selectable.
Alarm Output: SPST form C 110 VDC, 120 VDC (max.), 1 A (mas.), 30
W, 62.5 VA (resistive).
Dampening: (Measurement averaging) programmable from 0 to 16
consecutive readings.
Temperature Compensation: 20 to 120°F (-7 to 49°C).
Temperature Error: Maximum of ±0.004% of span per °F over
compensated range for zero and span.
Temperature Effect Electrical Measurement: ±0.001% of span per
°F over compensated range.
Ambient Operation Temperature: 32 to 120°F (0 to 49°C).
Storage Temperature: -4 to 158°F (-20 to 70°C).
Power Supply: Internal: two 9 VDC alkaline batteries,External: AC
adapter 9 VDC, 500 mA.
Battery Life: 30 hours (approximate).
Engineering Units: inH2O, psi, inHg, kpa, mbar, cmH2O,
mhG, and user-defined.
Display: Alphanumeric LCD, 0.37" (9.5 mm) height per line, 2
lines, 16 characters/line.
Display Update: 100 msec.
Process Connection: 1/8" Female NPT.
Electrical Connections: Miniature recessed banana jacks.
Housing: ABS plastic.
Weight: Calibrator: 2.2 lbs. (1.3 kg), Pressure Module: 0.5 lb.
(0.3 kg).
APPLICATIONS:
Pressure measurement for clean room control, filter performance
monitoring, HVAC testing and setup, draft measurement, airflow
measurement control, differential pressure measurement in laminar
flowhoods, point booths, industrial ovens and fume hoods. Use as a
secondary standard for calibrating pressure equipment.
|
 |
Model
CAL30 |
Process Calibrator
Measures and Generates DC Voltage, mV and Current,
Programmable. |
 |
|
Calibrate electronic process instruments such as controllers,
transmitters, or recorders with the model CAL30 Process Calibrator.
Model CAL30 is designed to independently measure and simulate DC volts,
mV, and current. These calibrators simulate a two-wire field transmitter
and generate a controlled output current when connected to an external
power supply. Units source is a four-wire transmitter by generating a
controlled 4-20 mA DC from an external 25 V power supply.
Programmable functions, such as square root extraction for flow
signals, ramp/step increment value, selectable engineering units,
isolated signal conversion, or automatic changing of the output value
simplify many calibration tasks. Calibration values can be stored and
recalled using the Low, Mid, or High keys. Program the automatic
shut-off to conserve battery life.
The Model CAL30 Calibrator features 500 VDC input/output isolation
and reverse polarity protection. The unit includes three 9V batteries,
two sets of dual test leads, a dual binding post adapter, carrying case,
and instruction manual.
POPULAR
MODELS
Model
Number |
Description |
Price |
| CAL30 |
Process Calibrator |
$1195.00 |
|
SPECIFICATIONS
Input Ranges: Voltage: -10 to 24 mV DC; -10 to 120 mV DC (autoranging);
0 to 12 VDC; Current: 0 to 24 mA DC
Accuracy: ±0.025% of range.
Output Ranges: 0 to 120.0 mV DC (2KΩ min. load); 0 to 12.0
VDC (2KΩ min. load); 0 to 24.0 mA DC (25 volts excitation)
Warm-up Time: 60 seconds, maximum.
Input Impedance: 1 MΩ minimum plus 10 nanoAmps maximum for V
and mV inputs; 40Ω for mA DC.
Input/Output Isolation: 500 VDC
Temperature Limits: -4 to 113°F (-20 to 45°C)
Display: 24-character LCD, 3/16" (5 mm) character height.
Indication: Up to 5 digits depending on range.
Electrical Connections: Standard banana jacks on ¾" (19 mm)
centers.
Power: Three 9V alkaline (included)
Battery Life: 100 hours, approx.
Weight: 1.15 lbs (0.52 kg)
|
 |
Model
CAL20 |
RTD Calibrator
Measures and Simulates Common RTDs, Programmable. |
 |
| Calibrate electronic process instruments such as
controllers, transmitters, transducers or recorders with the model CAL20
RTD Calibrator. Model CAL20 is designed to independently measure and
Simulate and common RTD signals (100Ω Pt, 120Ω Ni, 10Ω
Cu),ohms or measure current. Input/output can be operated simultaneously
with independent range selection.
A multitude of programmable functions, ramp/step increment value,
selectable engineering units, isolated signal conversion, or automatic
changing of the output value simplify many calibration tasks.
Calibration values can be stored and recalled using the Low, Mid, or
High keys. Program the automatic shut-off to conserve battery life.
The Model CAL20 Calibrator features 500 VDC input/output isolation
and reverse polarity protection. View input and output values on the 24-
character LCD. The unit includes three 9V batteries, two sets of dual
test leads, a dual binding post adapter, carrying case, and instruction
manual.
POPULAR
MODELS
Model
Number |
Description |
Price |
| CAL20 |
RTD Calibrator |
$1195.00 |
|
SPECIFICATIONS
Input Ranges: 100 Ω Pt: -364 to 1562°F (-220 to 850°C);
120 Ω Ni: -112 to 500°F (-80 to 260°C); 10 Ω Cu: -328 to 500°F
(-200 to 260°C).
Accuracy: ±0.025% of range.
Output Ranges: 100 Ω Pt: -364 to 1562°F (-220 to 850°C);
120 Ω Ni: -112 to 500°F (-80 to 260°C); 10 Ω Cu: -328 to 500°F
(-200 to 260°C).
Warm-up Time: 60 seconds, maximum.
Input/Output Isolation: 500 VDC.
Operating Temperature: -4 to 113°F (-20 to 45°C).
Display: 24-character LCD, 3/16" (5 mm) character height.
Indication: Up to 5 digits depending on range.
Electrical Connections: Standard banana jacks on ¾" (19 mm)
centers.
Power: Three 9V alkaline (included).
Battery Life: 100 hours, approx.
Weight: 1.15 lbs (0.52 kg)
|
 |
Model
LC21 |
Smart Loop Calibrator
Sources and Reads Milliamps, Auto-Stepping,
Auto-Ramping |
 |
|
Perform field calibration of 2-wire transmitters, receivers,
recorders, and controllers with Model LC21 Smart Loop Calibrator. The
portable LC21 calibrator sources 24 VDC, reads 0 to 21.00 milliamps and
simulates 2-wire transmitters. Unit features continuous output or fixed
step current output. Use fixed calibration currents to test and
calibrate equipment or to control valve stroking. Output currents can be
changed manually or automatically. The LC21 calibrator can source or
simulate continuous up/down cycling between 4 and 20 mA--ideal for
endurance testing. Display readings in mA or in % of 4-20mA or 0-20mA.
Unit includes carrying case, batteries, spare fuse, and manual.
POPULAR
MODELS
Model
Number |
Description |
Price |
| LC21 |
Smart Loop Calibrator |
$599.00 |
ACCESSORIES
Model
Number |
Description |
Price |
| LC21-110 |
110 VAC Adapter |
$45.00 |
| LC21-230 |
230 Vac Adapter |
$50.00 |
|
SPECIFICATIONS
Output Range: 0 to 21.00 mA.
Input Range: 0 to 21.00 mA.
Accuracy: ± 0.05% of range.
Resolution: 10 microamps.
Internal Loop Power: 24 VDC stabilized.
External Loop Power: 56 VDC maximum.
Fixed Steps: Output in mA
4-20mA lin: 4-8-12-16-20,
0-20mA lin: 0-5-10-15-20,
4-20mA sq.rt (flow): 4-5-8-13-20,
0-20mA sq.rt (flow): 0-1.25-5-11.25-20,
4-20mA valve: 3.8-4-4.2-12-19-20-21.
Fixed Step Interval: 10 seconds in auto-stepping mode.
Auto-Ramping Interval: 60 seconds between limits.
Temperature Effect: 0.0015% per °F, 0.003% per °C.
Drive Capability: 900 Ohms.
Temperature Limits: 14 to 122°F (-10 to 50°C), Storage: -65 to
260°F (-54 to 126°C).
Power: Four AA 1.5V alkaline batteries.
Battery Life: Sources 12 mA: 13 hours; Measures: 64 hrs.
Housing: High impact ABS.
Weight: 9oz (285g).
|
 |
Model
CA20 |
Thermocouple Calibrator
Accepts Type J Thermocouples, High/Low Alarm,
Selectable Resolution. |
 |
|
Dual function CA20 Thermocouple Calibrator can be used to
source Type J thermocouple output values or measure Type J
thermocouples. Quickly select output value using the rotary switch with
coarse and fine adjustments. Model CA20 accepts up to two Type J
thermocouples and features data hold and max functions. Measurements are
displayed in a 3½-digit backlit display. Units include a 4" (100
mm) Type J bead wire temperature probe, calibration cable, 9 V battery
and instruction manual.
POPULAR
MODELS
Model
Number |
Description |
Price |
| CA20 |
Thermocouple Calibrator |
$159.00 |
|
SPECIFICATIONS
Thermocouple Type: Type J.
Measurement Range: -40 to 1400°F (-40 to 760°C), selectable.
Calibration Range: -328 to 2192°F (-210 to 1200°C).
Accuracy: ±0.1% rdg + 2°F on -40 to 1400°F; ±0.1% rdg +1°C
on -40 to 760°C, over operating temperature of 64 to 82°F (18 to 28°C).
Resolution: Selectable 1° or 0.1 °.
Temperature Coefficient: 0.1 times the applicable accuracy
specification per °F from 32 to 64°F and 82 to 122°F (0 to 18°C and
28 to 50°C).
Reading Rate: 2.5 times/second.
Input Protection: 24 VDC or 24 VAC rms max input voltage on any
combination of input pins.
Operating Temperature: 32 to 122°F (0 to 50°C) max 80% RH.
Storage Temperature: -4 to 140°F (-20 of 60°C) max 70% RH.
Display: 0.8" (20 mm) height, 3½-digit LCD with switchable
back light.
Thermocouple Connection: Standard female mini-connector.
Battery: 9V Alkaline (Included).
Battery Life: 200 hours typical.
Housing: ABS Plastic.
Weight: 15.5 oz. (440 g).
|
 |
Model
CA10 |
Thermocouple Calibrator
Accepts Type K Thermocouples, High/Low Alarm, Selectable
Resolution. |
 |
|
Dual function CA10 Thermocouple Calibrator can be used to
source Type K thermocouple output values or measure Type K
thermocouples. Quickly select output value using the rotary switch with
coarse and fine adjustments. Model CA10 accepts up to two Type K
thermocouples and features data hold and max functions. Measurements are
displayed on a 3½-digit backlit display. Units include a 4" (100
m) Type K bead wire temperature probe, calibration cable, 9 V battery
and instruction manual.
|
Model
Number |
Description |
Price |
| CA10 |
Thermocouple Calibrator |
$1195.00 |
 |
Model
CAL10 |
Multi-Cal Benchtop Calibrator
Measures Thermocouples, Voltage and Current,
Programmable |
 |
|
Calibrate electronic process instruments such as controllers,
transmitters, transducers or recorders with Model CAL10 Thermocouple
Calibrator. Model CAL10 is designed to independently measure and
simulate dc mV or V and common thermocouple signals (Type J, K, T, E, J
(DIN), and T (DIN) or measure current. Input/output can be operated
simulanteously with independent range selection.
A multitude of programmable functions, such as square root extraction
for flow signals, ramp/step increment value, selectable engineering
units, isolated signal conversion or automatic changing of the output
value simplify many calibration tasks. Calibration values can be stored
and recalled using the Low, Mid, or High keys. Cold junction
compensation can be selected for automatic or manual. Program the
automatic shut-off to conserve battery life.
The Model CAL10 calibrator features 500 VDC input/output isolation
and reverse polarity protection. The unit includes three 9V batteries,
two sets of dual test leads, a dual binding post adapter, carrying case,
and instruction manual.
POPULAR
MODELS
Model
Number |
Description |
Price |
| CAL10 |
Thermocouple Calibrator |
$1195.00 |
|
SPECIFICATIONS
Input Ranges: Voltage: -10 to 24 mV DC/-10 to 120 mV DC (autoranging),
0 to 12 VDC; Current: 0 to 24 mA DC; Temperature: see table below.
Accuracy: Voltage and Current: ±0.025% of range; Temperature:
see table below.
Output Ranges: 0 to 120 mV DC (2K Ω min. load); 0 to 12 VDC
(2K Ω min. load); Thermocouple Types J, K, T, E, J (DIN), and T
(DIN) similiar to input ranges.
Resolution: 0.1°C/F.
Response Time: 5 seconds.
Temperature Limits: -4 to 113°F (-20 to 45°C).
Display: 24-character LCD, 3/16" (5 mm) character height.
Indication: Up to 5 digits depending on range.
Electrical Connections: Standard banana jacks on 3/4" (19
mm) centers.
Power: Three 9V alkaline (included).
Battery Life: 100 hours, approximate.
Weight: 1.15 lbs. (0.52 kg).
| T/C |
Range |
Accuracy |
| J |
-250 to 2192°F
346 to -250°F |
±1°F
±2°F |
| K |
-350 to 2500°F
-380 to -350°F
-420 to -380°F
| ±1°F
±3°F
±5°F |
| T |
-100 to 1112°F
-300 to -100°F
-400 to -300°F |
±1°F
±3°F
±5°F |
| E |
-400 to -1832°F
-420 to -400°F |
±1°F
±2°F |
|
|