HI 991300 • HI 991301
Waterproof pH, EC, TDS & Temperature Meters
- 4 instruments in one! - pH, conductivity, TDS and
temperature all in one meter with a single probe.
- Waterproof - HI 991300 and HI 991301 are housed in a
compact waterproof, IP67 rated case.
- Tailor made for your application - Customize pH/EC
calibration, select EC/TDS conversion factor and temperature
coefficient and choose the temperature scale that best fits your
application.
- Ease of use, accuracy and durability - Two button
operation, microprocessor based, rugged instrument and probe
construction.
- Order
Information
- Specifications
|
|
These new waterproof instruments offer pH, conductivity, TDS (Total Dissolved
Solids) and temperature readings in one easy to use meter. HI 991300
measures low range conductivity and TDS, while HI 991301 measures high
range. Both instruments provide full scale pH readings and allow the user to
select the temperature readout in Centigrade or Fahrenheit. Other features
include automatic calibration, automatic temperature compensation, selectable
EC/TDS conversion factor, adjustable conductivity temperature coefficient,
stability indicator, Battery Error Prevention System and battery strength
percentage guage. All functions are accomplished with only two buttons. When the
instrument is turned on, % battery power remaining is displayed preventing
unexpected battery failures in the field. Calibration is quick and simple at one
or two points and the user can select between the instruments’ memorized NIST
or standard pH buffers. Conductivity and TDS calibration is also automatic with
five programmed values to choose from. Depending on the application, the user
can select the conversion factor between conductivity and TDS, as well as the
temperature coefficient factor. All calibration and user settings are saved even
during battery replacement. The HI 1288 multiparameter probe
compliments the meters’ outstanding features: the pH sensor utilizes gel
electrolyte and a cloth junction for longer life, faster response and greater
durability. The conductivity/TDS sensor is composed of graphite. This new
technology eliminates oxidation and polarization errors which are the main
causes for erroneous readings and frequent calibration. For added protection,
the probe is housed in a rugged plastic sheath with reinforced rubber cable and
steel DIN connector.
HI 991300 is supplied with pH/EC/TDS/°C probe, batteries and
instructions in a hard carrying case.
HI 991301 is supplied with pH/EC/TDS/°C probe, batteries and
instructions in a hard carrying case.
| Accessories |
| HI 1288 pH/EC/TDS/°C probe |
| HI 710004 Soft carrying case |
| HI 70031P 1413 µS/cm calibration packets (20 mL;
25 pcs): (HI 991300) |
| HI 70030P 12880 µS/cm calibration packets (20 mL;
25 pcs): (HI 991301) |
| HI 70038P 6.44 ppt cal. sol. (25 x 20 mL) |
| HI 77400P pH 7 and 4 calibration packets (20 mL; 5
pcs each) |
| HI 77100P 1413 µS & pH 7 sol., 20 mL, 10 pcs
each |
| HI 77200P 1500 ppm & pH 7 sol., 20 mL, 10 pcs
each |
| HI 77300P 1382 ppm & pH 7 sol., 20 mL, 10 pcs
each |
| HI 77400P pH 4 & 7 sol., 20 mL, 5 pcs each |
| HI 710007 Shockproof rubber boot (blue) |
| HI 710008 Shockproof rubber boot (orange) |
| Specifications: |
HI 991300 |
HI 991301 |
| Range |
|
|
| pH |
0.00 to 14.00 pH |
0.00 to 14.00 pH |
| EC |
0 to 3999 µS/cm |
0.00 to 19.99 mS/cm |
| TDS |
0 to 2000 ppm |
0.00 to 10.00 ppt |
| ºC/ºF |
0.0 to 60.0°C (32 to 140°F) |
0.0 to 60.0°C (32 to 140°F) |
| Resolution |
|
|
| pH |
0.01 pH |
0.01 pH |
| EC |
1 µS/cm |
1 mS/cm |
| TDS |
1 ppm |
1 ppt |
| ºC/ºF |
0.1ºC/ºF |
0.1ºC/ºF |
| Accuracy (@20°C) |
|
|
| pH |
±0.01 pH |
±0.01 pH |
| EC/TDS |
2% Full Scale |
2% Full Scale |
| ºC/ºF |
±0.5°C (± 1.0ºF) |
±0.5°C (± 1.0ºF) |
| Typical |
|
|
| EMC Deviation |
pH |
±0.3 pH |
±0.3 pH |
| EC/TDS |
2% Full Scale |
2% Full Scale |
| ºC/ºF |
±1°C (± 1.0ºF) |
±1°C (± 1.0ºF) |
|
| EC/TDS Calibration |
1 point Automatic
at 1382 ppm (conv.=0.5);
1500 ppm (conv.=0.7);
1413 µS/cm (others) |
1 point Automatic
at 6.44 ppt (conv.=0.5);
9.02 ppt (conv.=0.7);
12880 µS/cm (others) |
| pH Calibration |
Automatic 1
or 2 points with 2 sets of memorized buffers
(4.01/7.01/10.01 or 4.01/6.86/9.18) |
| EC/TDS Ratios |
Selectable
between 0.45, 0.50 (default),
0.55, 0.60, 0.65, 0.68, 0.70, 0.75, 1.00 ppm=1 µS
|
Temperature
Compensation |
pH: Automatic
from 0 to 60°C;
EC/TDS: Automatic from 0 to 60°C with selectable B
of 0.0, 1.8, 1.9 (default), 2.0, 2.1, 2.2, 2.3, 2.4% per °C
|
| Probe |
HI 1288 pH/EC/TDS
with temp sensor, DIN connector and 1 m (3.3’) cable (included) |
| Battery Life/Type |
500 hours of continuous use/4 x 1.5V
AAA |
| Environment |
0 to 50°C (32 to 122 °F); RH 100% |
| Dimensions |
143 x 80 x 38 mm (5.6 x 3.1 x
1.5’) |
| Weight |