pHep® - Waterproof pH Testers with Replaceable Electrode
Cartridge
- Dual Level Display Showing pH & Temp.
- Percent Remaining Battery Indicator.
- Replaceable pH Electrode Featuring an Extendable Junction.
- Rugged, Floating Waterproof Casing.
- Automatic Calibration at standard as well as NIST Buffer Sets.
- Measurement Stability Indicator.
- Order
Information
- Specifications
|
|
When Hanna Instruments introduced it's first version of the pHep (pH electronic
paper) in 1986, it immediately revolutionized the standards in pH
measurement. The pHep's ease-of-use in combination with its affordability made
it possible for non-technical and technical personnel alike to accurately
measure pH on the spot. Since then, over 1 million pHeps have gone to satisfied
customers around the world. We have always valued your comments and suggestions
and in this new series, we have included all the features you have asked for and
more!
The waterproof pHep® 4 and pHep® 5 pH
testers have a dual-level LCD that displays pH and °C or °F. At startup, pHep®
performs a self check then displays the remaining battery level to
assure proper working condition. These testers also feature a stability
indicator and hold feature to prompt the user when to take the reading &
freeze the display for easy and accurate recording. These new pH/Temperature
meters have a stainless steel temperature probe for faster and more precise
temperature measurement and auto-compensation. 1 or 2 point calibration is made
simple with automatic recognition of both standard and NIST buffer sets. We have
designed the new pHep® pH electrodes with an extendable
reference junction and when the electrode finally is exhausted simply insert a
new electrode cartridge!
 |
Replaceable electrode cartridge
featuring an extendable junction. |
- pHep® 4 (HI 98127) has the most
common pH range of 0.0 - 14.0 pH
- pHep® 5 (HI 98128) has a more
precise pH range of 0.00 - 14.00 pH
pHep® is supplied complete with 4 x 1.5V batteries, pH
electrode (HI 73127) and electrode replacement tool (HI 73128).
| Specifications: |
|
pHep® 4
(HI 98127) |
pHep® 5
(HI 98128) |
| Range |
pH |
0.0 to 14.0 pH |
0.00 to 14.00 pH |
| Temperature |
0.0 to 60.0 °C or 32.0 to 140.0 °F |
| Resolution |
pH |
0.1 pH |
0.01 pH |
| |
Temperature |
0.1 °C or 0.1 °F |
| Accuracy (@ 20°C/68°F) |
pH |
±0.1 pH |
±0.01 pH |
| |
Temperature |
±0.5 °C or ±1 °F |
| Typical EMC Deviation |
pH |
±0.02 pH |
| |
Temperature |
±0.5 °C or ±1 °F |
| Calibration |
|
1 or 2 points with 2 sets of memorized
buffers:
(pH 4.01, 7.01 & 10.01 or 4.01, 8.68 & 9.18) |
Temperature
Compensation |
|
Automatic |
| Battery Type / Life |
|
4 x 1.5V / 350 hours |
| Environment |
|
0 to 50°C (32 to 122°F);
R.H. 100% |
| Dimensions |
|
163 x 40 x 26 mm (6.4 x 1.6 x 1.0") |
| Weight |
|
85 g (3 oz.) |
|