The Model 3330 Laboratory PH/Ion Meter |
|
Dual Display Simultaneous display of
primary and secondary measurements. A Switch Screen key is used to toggles
the input channel priority.
Ease of Use A simple push-button control panel, used in conjunction
with the display, ensures that even inexperienced operators can achieve
reliable results.
Dual Input (Rear Panel) The Model 3330 Laboratory PH/Ion Meter
features two independent input channels for both sensor and temperature
monitoring. A socket is provided for connection to an optional overhead or
bench stirrer which can then be controlled directly from the instrument
keypad.
Set Up Menus Set-up menus are provided on the Model 3330
Laboratory PH/Ion Meter to give access to all user defined options.
|
Languages Clear user instructions are available in six language
options, (English, French, German, Italian, Portuguese or Spanish),
operator selectable via the instrument set up mode.
Calibration A choice of 1 to 3 point calibration is offered for pH
measurement and 1 to 5 point calibration in Ion mode. The Model 3330 will
automatically recognize 2, 4, 7, 9 and 10pH buffers with a manual override
facility.
Sample Measurement Icons and prompts guide the user through
calibration and measurement routines.
GLP Essential to the modern laboratory GLP protocols are fully
supported. Product performance is continually monitored, with all logged
results being time and date stamped, thus satisfying the requirements of
Good Laboratory Practice.
RS232 / Analog Outputs The Model 3330 Laboratory PH/Ion Meter can
download directly to a computer or printer by the bi-directional RS232
link and provides an analogue output for connection to a chart recorder or
datalogger.
Other Ion Meters
available from Techne® are the Model
3205 Portable Ion Meter
and the Model 3345
Laboratory Ion Meter.
See
our available Electrochemistry accessories.

Model 3330 Specifications |
| pH |
|
| Range |
-2 to 19.999pH |
| Resolution |
0.1 / 0.01 / 0.001pH |
| Accuracy |
±0.003pH |
| mV |
|
| Range |
0 to 1999.9mV |
| Resolution |
0.1mV |
| Accuracy |
±0.2mV |
| Concentration |
|
| Range |
1 x 10-9 to 9.99 x 109 |
| Units |
Activity, ppm, %, mg/l, M or none |
| Resolution |
3 significant digits |
| Accuracy |
±0.003pX for monovalent ions |
| Measurement Methods |
Direct potentiometry |
| Input Impedance |
>1013 Ohms |
| Number of Electrode Inputs |
2 |
| Temperature |
|
| Range |
-10 to 105ºC / 14 to 221ºF |
| Resolution |
0.1ºC / 1ºF |
| Accuracy |
±0.5ºC / ±1ºF |
| ATC Range |
0 to 100ºC / 32 to 212ºF |
| Manual Temperature Compensation Range |
0 to 100ºC / 32 to 212ºF |
| Number of Temperature Inputs |
2 |
| Auto Buffer Recognition |
4, 7, 9.22 and 10.05 pH buffers with manual override facility |
| Calibration |
1, 2 or 3 point pH / 1, 2, 3, 4 or 5 point Concentration |
| Outputs |
Analog 1mV per digit / Bi-directional RS232 / hi-lo alarm
outputs open collector 0.5A 50V max |
| Datalogger |
100 results |
| Clock |
24 hour, hour/min/sec or day of month/month/year - Leap year
corrected |
| Alarm Points |
-2 to 19.999pH / ±1999.9mV / 1 x 10-9 to 9.99 x
109 |
| Good Laboratory Practice |
Calibration reminder with an interval of 1 to 999 hours from
last calibration. Timed printout with an interval from 1 second to 1
day |
| Display |
LCD Graphics |
| Power |
Power supply 9V AC |
| Size |
275(l) x 240(w) x 150(d)mm |
| Weight |
1.2Kg |

|
558 101 |
Model 3340 Laboratory Ion Meter supplied with glass
bodied combination electrode (924 005), swing arm electrode holder
(903 200), ATC probe (027 232), English instruction manual, 4 &
7 pH buffer capsules, US 115 volt power supply and BNC to Leadless
electrode cable (013 173) |
|
027 232 |
Temperature Probe |
|
013 173 |
Electrode lead (BNC to Leadless) |

|