+GF+ Signet Vortex Systems
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
+GF+ SIGNET Vortex Systems
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Operating Range: | d20 - d25 (0.5 - 0.75
in.) sensors: 0.5 to 4 m/s (1.6 to 13 ft/s) d32 - d40 (1.0 - 1.25 in.) sensors: 0.4 to 4 m/s (1.3 to 13 ft/s) d50 - d63 (1.5 - 2.0 in.) sensors: 0.3 to 4 m/s (1.0 to 13 ft/s) |
| Pipe Size Range: | d20mm - d63 mm, DN15mm- DN50 mm, 0.5 in. - 2 in. |
| Accuracy: | +/-1% of reading @ 25 degrees C |
| Repeatability: | +/-0.5% of reading @ 25 degrees C |
| Display - System Version Only: | Type: 8-digit alphanumeric
dot matrix LCD Contrast: Automatic, temp. compensated Update rate: 1s |
| Totalizers - System Version Only: | 8-digit resettable with
security option 8-digit non-resettable |
| Materials: | Sensor: PVC, PP, PVDF, or
HP-PVDF Union O-Rings: FPM or EPDM Enclosure: Glass-filled polypropylene, NEMA 4X/IP65 Seals (2): BUNA-N Window - System version: SMMA (Novacor) |
| Immunity: | EN50082-2 |
| Emissions: | EN55011 Reverse polarity protection |
| Quality Standards: | CE, Manufactured under ISO 9001 |
| Power: | 3.5 to 24 VDC, regulated, 1.5 mA max |
| Output type: | Open-collector NPN
transistor, 10 mA max sink, 24 VDC max pull-up voltage, 0 to 400 Hz (size-dependant), 50%
duty cycle, non-isolated
|
| Power: | 7 to 30 VDC, regulated, 20 mA max |
| Current output: | Loop impedance: 1 max at 7 VDC, 300 max at 12 VDC, 800 max at 24 VDC, 1000 max at 30 VDC |
| Resolution: | 6 µA |
| Current loop/system power: | 17 to 30 VDC, regulated, 23 mA max |
| Current loop: | Reversible span |
| Loop impedance: | 1 max at 17 VDC, 300 max at 24 VDC, 800 max at 30 VDC |
| Update rate: | <100 milliseconds |
| Resolution: | 6 µA |
| Sensor pulse output: | Open-collector
transistor, sink or source capability, optically isolated, 10 mA max sink, 30 VDC max
pull-up voltage, 0 to 400 Hz (size dependant), 50% duty cycle
|
| Maximum Media PVDF: | 16 bar @ 0 degrees C, 6.5 bar @ 100 degrees C |
| Pressure/Temperature: | (232 psi @ 32 degrees F, 94
psi @ 212 degrees F) PP: 10 bar @ 0 degrees C, 1.5 bar @ 80 degrees C (145 psi @ 32 degrees F, 22 psi @ 176 degrees F) PVC: 16.2 bar @ 0 degrees C, 3.7 bar @ 60 degrees C (235 psi @ 32 degrees F. 54 psi @ 140 degrees F) |
| Ambient temperature: | 0 to 70 degrees C (32 to 158 degrees F) |
| Storage temperature: | -15 to 80 degrees C (5 to 176 degrees F) |
| Relative humidity: | 0 to 95%, non-condensing |
| Maximum vibration: | 1 mm or 1g double amplitude
@ 500 Hz
|
|
![]() |
| A Reynolds number is a parameter describing flow velocity, viscosity, and pipeline size. To maintain system accuracy, a Reynolds number >10000 is required. To calculate the Reynolds number, use the following formula: | Reynolds
number = DV/u D = internal pipe diameter (m) V = flow velocity (m/s) u = kinematic viscosity (m2/s) Water (u) = 1·10-6m2/s |
Minimum downstream pipe backpressure levels are required to prevent cavitation within the sensor. The minimum back pressure is calculated by the following formula: 2.7P + 1.3 x Po (P = Pressure drop across sensor. Po = Water saturation vapor pressure at operating temperature.)
|
Pressure Drop Graph |
![]()
|
![]() |

![]()
|
![]() |

![]()
|
||||||||||||||||||||
|
||||||||||||||||||||
|