Techne Temperature Calibration Equipment
Precision FB08 Series Fluidised
Baths
Description and operation The FB08 is a rectangular bench standing unit. The controls for temperature
and fluidizing air flow are mounted on a recessed panel on the front of the
unit, together with on-off switches for the power supply and the dust extractor
and indicators for various functions. Insulated and vented the outer case is
safe to touch even at the maximum operating temperature. The inner container is
filled from the top with alumina. When fluidised, this medium is heated by four
immersion heaters close to the container wall with the control thermocouple
between the heaters. The internal components are protected by an air cutout
switch that disables heat in the event the fluidizing air is
lost. In operation, set the fluidizing air by means of a combined control valve and flow meter. (Note: air adjustment is automatic with the FB08C) Fluidizing air requirements are inverse of the bath temperature. i.e- lower bath temperatures require more air for proper fluidization. The bath temperature is set by means of a front panel mounted PID controller. Sensors for calibration should be suspended as close as possible to the center of the bath. When calibrating sensors the probe holder is recommended as it is custom built with the dimensions of the sensors you will be calibrating. Dead-Bed Operation
Models to choose from: The FB08 is offered in three different models. The standard FB08 operates from +50 to 700°C, in addition to the standard unit the FB08C offers automatic fluidizing air control and is capable of remote PC operation. When connected to your liquid nitrogen supply the FB08LT operates between -100 and +200°C. The single probe plate and ring are included with the baths. You will also receive one 35lb. charge of alumina to get started. Specifications
Accessories
TC-8D Electronic Control Unit -- $1,552
Laboratory Fluidised Baths Five standard models offer a wide choice of bath sizes and temperature ranges.
Specifications:
Ordering Information:
LCB-7+12 Series Calibration Baths --
$1,243
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
LCB-7 |
LCB-7 Calibration Bath Compact, accurate and reliable liquid baths for calibration.
|
| Controllers incorporate the safety features
of overtemperature cut-out and low liquid level device. LCB-7 baths have a
thermometer hole to take a reference thermometer, drain tap, lid, carrying
handles and insulation on sides and base. A built in cooling coil is
included. The system offers a display accuracy of ±1.0°C, but if you
require higher accuracy take a look at one of the thermometer
systems we offer. We recommend the use of our bath liquids to get the best performance at a wider range of operating temperatures. |
General Features:
| LCB-7D** | LCB-7C** | LCB-7HT | |
| Bath Capacity | 7 liters | 7 liters | 7 liters |
| Operating temperature range1 | -35 to 200 | -35 to 250 | NA |
| Working temperature range2 | 25 to 200 | 50 to 250 | 25 to 250 |
| Display resolution | 0.1 | 0.1 | 0.1 |
| Stability at 50°C | +/-0.01 | +/-0.01 | +/-0.01 |
| Accuracy | +/-1.0 | +/-1.0 | +/-1.0 |
| Temp. Sensor | PRT | PRT | PRT |
| Working Area in. LxWxD | 5.6x6x7 | 5.6x6x7 | 5.6x6x7 |
| Controller length/width | 9.3/4.9 | 9.3/4.9 | 9.3/4.9 |
| Height with controller | 13.5 | 13.5 | 13.5 |
1) With an FC-500 flow cooler; 2) ambient of 20°C
Ordering Information:
| Catalog Number | Description: |
| 3022300 | LCB-7 bath less controller |
| 1016300 | **TU20D - 117 volt |
| 1016700 | **TU20D - 240 volt |
| 1016400 | **TU20C - 117 volt |
| 1016900 | **TU20C - 240 volt |
| 1017600 | TU20HT - 117 volt |
| 1017650 | TU20HT - 240 volt |
**For below ambient temperatures inquire about Techne's flow
coolers.**
|
LCB-12 |
LCB-12 Calibration Bath Compact, accurate and reliable liquid baths for calibration.
|
The LCB-12 bath has a thermometer hole to take a reference
thermometer, drain tap, lid, carrying handles and insulation on sides and base.
A built in cooling coil is included. The 12 inch immersion depth of the LCB-12
makes it an excellent choice for use in the calibration of liquid in glass
thermometers, RTD's, thermocouples and fluid filled bulbs. Controllers
incorporate the safety features of an over-temperature cut-out and low liquid
level device. The system offers a display accuracy of ±1.0°C, but if you require
higher accuracy take a look at one of the thermometer
systems we offer.
We recommend the use of
our bath liquids
to get the best performance at a wider range of
operating temperatures.
General features:
| LCB-12D** | LCB-12C** | LCB-12HT | |
| Bath Capacity | 12 liters | 12 liters | 12 liters |
| Operating temperature range1 | -35 to 200 | -35 to 250 | NA |
| Working temperature range2 | 25 to 200 | 50 to 250 | 25 to 250 |
| Display resolution | 0.1 | 0.1 | 0.1 |
| Stability at 50°C | +/-0.01 | +/-0.01 | +/-0.01 |
| Accuracy | +/-1.0 | +/-1.0 | +/-1.0 |
| Temp. Sensor | PRT | PRT | PRT |
| Working Area in. LxWxD | 5.5x5.75x12 | 5.5x5.75x12 | 5.5x5.75x12 |
| Controller length/width | 9.3/4.9 | 9.3/4.9 | 9.3/4.9 |
| Height with controller | 18.5 | 18.5 | 18.5 |
1) With an FC-500 flow cooler; 2) ambient of 20°C
Ordering Information:
| Catalog Number | Description: |
| 3022400 | LCB-12 less controller |
| 1016300 | **TU20D - 117 volt |
| 1016700 | **TU20D - 240 volt |
| 1016400 | **TU20C - 117 volt |
| 1016900 | **TU20C - 240 volt |
| 1017600 | TU20HT - 117 volt |
| 1017650 | TU20HT - 240 volt |
**For below ambient temperatures inquire about Techne's flow coolers.**
|
Portable
Thermoregulators | |
| Inventors of the "Clip On" thermoregulator in 1948, Techne® now offers five "Clip On" units. These Thermoregulators will convert your existing liquid bath or container into a precision temperature controlled bath with exceptional thermal performance. They offer a full range of options such as analog or digital set/readout, high pump output and RS232 interface. | |
|
|
The
TE10A clip on thermoregulator offers a temperature range
capability of -20°C to + 95°C, covering most
routine laboratory applications. A circulating pump with external
circulation connections is standard, as is adjustable over-temperature
cutout. |
|
|
The TE10D
Tempunit offers a temperature range capability of between
-40°C to +120°C. There is a digital temperature
selector, circulating pump with external circulation connections,
adjustable over-temperature cut-out and low liquid level cut-out. The
TU20D has a temperature range capability of -40°C to
+200°C and incorporates the additional feature of an RS232
port. The top of the range TU20C has additional features fitted as
standard including an audible alarm, a high capacity dual pressure pump,
cooling coil and also includes an RS232 port. The temperature range
capability is -40°C to +250°C.The TU20HT is designed for internally circulated baths, thus it is not capable of external pumping for re-circulation, but does include the cooling coil, Thermsoft software and has an operating temperature range capability of -40 to +250 °C. For applications that
require sub-ambient operation, see our refrigeration
units for use in combination with the unheated baths or one of the
refrigerated
baths for an integrated system.Download the full version of ThermSoft software. It allows you to create ramp and dwell programs and remotely control the TU20D, TU20C and TU20HT from your PC . Thermsoft is included with the TU20C and TU20HT and is a free download for the TU20D. |
Techne® Thermoregulator Specifications
|
Model |
TE-10A |
TE-10D |
TU-20D |
TU-20C |
TU-20HT |
|
Temperature range °C* |
*-20 to +95 | *-40 to +120 | *-40 to +200 | *-40 to +250 | *-40 to +250 |
|
Temperature selection |
analogue |
digital |
digital |
digital |
digital |
|
Temperature stability using water @40 °C |
+/- 0.01 |
+/- 0.01 |
+/- 0.005 |
+/- 0.005 |
+/- 0.005 |
|
Accuracy |
+/-2.0 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
+/-1.0 |
|
Pump capacity (liters/min) |
10 |
10 |
10 |
20/15** |
Internal 10 LPM
only |
|
Pump capacity (mbar) |
145 |
145 |
145 |
450 |
Internal 145mbar
only |
|
Method of control |
Proportional |
PID |
PID |
PID |
PID |
|
Temperature sensor |
Thermistor |
PRT |
PRT |
PRT |
PRT |
|
Adjustable over temperature cut out |
Yes |
Yes |
Yes |
Yes plus additional audible alarm |
Yes |
|
Low liquid level cut-out |
No |
Yes |
Yes |
Yes |
Yes |
|
PC interface |
No |
No |
Yes RS232 |
Yes RS232 |
Yes RS232 |
|
Cooling coil*** |
No |
No |
No |
Yes |
Yes |
|
Dimensions (mm) |
145 |
145 |
145 |
145 |
145 |
|
Overall size |
237 x 124 x 260 |
237 x 124 x 260 |
237 x 124 x 260 |
237 x 124 x 260 |
237 x 124 x 260 |
| * Minimum operating temperature is 5°C above your ambient. The
TU20C minimum is ambient plus 20°C. Refrigeration
required for below ambient cooling. ** 20 litres/minute pressure, 15 litres/minute suction. *** Available as an option where not supported as standard. |
|||||
Ordering Information:
| Model | Catalog # | Voltage | HZ | Watts | Shipping Wght lbs. |
| TE10A | 1016600 | 240 | 50/60 | 1000 | 8.1 |
| TE10A | 1016100 | 120 | 50/60 | 1000 | 8.1 |
| TE10D | 1016700 | 240 | 50/60 | 1000 | 8.5 |
| TE10D | 1016200 | 120 | 50/60 | 1000 | 8.5 |
| TU20D | 1016800 | 240 | 50/60 | 1500 | 8.8 |
| TU20D | 1016300 | 120 | 50/60 | 1500 | 8.8 |
| TU20C | 1016900 | 240 | 50/60 | 1800 | 9.2 |
| TU20C | 1016400 | 120 | 50/60 | 1800 | 9.2 |
| TU20HT | 1017650 | 240 | 50/60 | 1800 | 8.8 |
| TU20HT | 1017600 | 120 | 50/60 | 1800 | 8.8 |
REFRIGERATED BATHS
Download the
ESRB11 manual (1.0mb pdf)
![]() |
There are two different models of refrigerated
bath to choose from. These baths are a complete refrigerated circulating
system for open or closed applications for temperature ranges from
-25
|
|
SPECIFICATIONS - Refrigeration Units | ||
|
RB-5A |
ESRB-11 | |
Minimum achievable temperature (°C) |
-20 | -25 |
***Cooling capacity at 0°C (W) |
145 | 235 |
Cooling capacity at 20°C (W) |
145 | 371 |
Cooling capacity at -10°C (W) |
110 | 154 |
| Dimensions | ||
| *Overall size L x W x H in. | 16.9x9.2x17.5 | 20x15.2x25 |
| Liquid surface to top of bath - max (in.) | 2.5 | 2.5 |
| Working area - L x W x H in. | 7.5x5.9x7.8 | 7.7x11.5x5.9 |
| Working depth - Max/Min (in.) | 7.0/5.3 | 5.1/3.9 |
| Working capacity - Max/Min (liters) | 7.0/5.5 | 11.6/9.6 |
| **Shipping weight (lb) | 68 | 99 |
| Catalog number (120 volt) | 1071400 | 1071300 |
*Overall size includes a mounted TE10 or TU20 model
Thermoregulator.
**Add controller and bath
shipping weight o get shipping weight of complete bath
system.
***Using Techne® Catalog #7012602 bath
fluid
REFRIGERATED BATHS
![]() |
RB-5A/ESRB-11 There are two different models of refrigerated bath to choose from.
These baths are a complete refrigerated circulating system for open or
closed applications for temperature ranges from -25
|
|
RB-5A |
ESRB-11 | |
Minimum achievable temperature (°C) |
-20 | -25 |
***Cooling capacity at 0°C (W) |
145 | 235 |
Cooling capacity at 20°C (W) |
145 | 371 |
Cooling capacity at -10°C (W) |
110 | 154 |
| Dimensions | ||
| *Overall size L x W x H in. | 16.9x9.2x17.5 | 20x15.2x25 |
| Liquid surface to top of bath - max (in.) | 2.5 | 2.5 |
| Working area - L x W x H in. | 7.5x5.9x7.8 | 7.7x11.5x5.9 |
| Working depth - Max/Min (in.) | 7.0/5.3 | 5.1/3.9 |
| Working capacity - Max/Min (liters) | 7.0/5.5 | 11.6/9.6 |
| **Shipping weight (lb) | 68 | 99 |
| Catalog number (120 volt) | 1071400 | 1071300 |
* Overall size 6.7 x 11.0 x 10.2"
* Stability: ±0.05°C
* Display accuracy: ±0.5°C
* Display resolution: 0.1š
* Cools to a minimum temperature in 20 minutes
* Carrying case included
* Immersion depth for probes: 4" (single well)
* Deep version also available: 8" immersion (DB-35LD)
* Switchable main power 110V or 220V or battery operation

Click here to
learn how we heat and cool the Tecal 140.
Need a lower operating temperature? Connect
a cold water supply to the
Tecal 140 heatsink and achieve a temperature
45 degrees C. below the water temperature.
|
Tecal 140 Technical data | |||
|
Minimum temperature |
ambient -45 °C
(1) |
Heating rate 20 °C to
100°C |
5 min |
|
Maximum temperature |
140 °C |
Cooling rate 100 °C to
0°C |
9 min |
|
Temperature accuracy in measuring zone |
+/0.3 °C |
Programmable ramp rate °C/min |
0.1 to 25 on H version |
|
Temperature uniformity in measuring zone |
+/-0.2 °C
(2) |
Switch test |
On H version |
|
Measuring zone |
0 to 50mm from base of well |
Comms port, 9 way D-type |
Full bi-directional RS-232 |
|
Temperature stability after 10 mins |
+/-0.05 °C |
Weight (lbs) |
30 |
|
Display resolution |
0.1 °C or
0.1°F |
Dimensions |
285 x 190 x 426 |
|
1 -40° C absolute minimum with 5° C water circulated through block.2 At 100 °C | |||
|
Catalog # |
Model |
Voltage |
Hz |
Watts |
Wght lbs. |
|
3026500 |
Tecal 140S |
230V |
50/60 |
400 |
29.8 |
|
3025900 |
Tecal 140S |
120V |
50/60 |
400 |
29.8 |
|
3026600 |
Tecal 140H |
230V |
50/60 |
400 |
29.8 |
|
3026000 |
Tecal 140H |
120V |
50/60 |
400 |
29.8 |
|
Catalog Number |
Description |
Weight |
|
7032533 |
Insert 5 x ¼" probes |
13oz |
|
7032534 |
Insert 3/8" + 5/16" +1/4" + 3/16" + 1/8" probe |
13oz |
|
7032535 |
Insert 2 x 3/8" + 2 x ¼" probes |
13oz |
|
7032536 |
Insert 2 x ½" + 2 x ¼" probes |
13oz |
|
7032537 |
Insert 1 x ¼" probes |
13oz |
|
7032538 |
Insert blank |
13oz |
|
7032574 |
Insert 9/16" probe |
13oz |
|
7032575 |
Insert 5/8" probe |
13oz |
|
7032576 |
Insert 11/16" probe |
13oz |
|
7032577 |
Insert ¾" probe |
13oz |
Accessories
|
Catalog Number |
Item |
Voltage |
HZ |
Watts |
Weight lbs. |
|
7032578 |
Software diskette and RS232 cable |
||||
|
2040400 |
CH-5 chiller unit - contact us for more info on this item. |
120V |
50/60 |
1200 |
58.4 |
Tecal 425 S/H
Mid Temperature Range
Block
Calibrator

|
Tecal 425 Technical data | |||
|
Minimum temperature above ambient |
20 °C Above
ambient |
Cooling rate 400 °C to
100°C |
25 min |
|
Maximum temperature |
425 °C |
Programmable ramp rate °C/min |
0.1 to 25 on H version |
|
Temperature accuracy in measuring zone |
+/0.3 °C |
¼" reference hole in block |
Yes |
|
Temperature uniformity in measuring zone |
+/-0.2 °C1 |
Switch test |
On H version |
|
Measuring zone |
0 to 50mm from base of well |
Fan cooling |
Automatic |
|
Temperature stability after 10 mins |
+/-0.05 °C |
Comms port, 9 way D-type |
Full bi-directional RS232 |
|
Display resolution |
0.1 °C or
0.1°F |
Weight (lbs) |
20.0 |
|
Heating rate, 20 °C to
400°C |
15 min |
Dimensions |
285 x 190 x 426 |
|
1 At 400°C | |||
Ordering Information
|
Catalog # |
Model |
Voltage |
Hz |
Watts |
Wght lbs. |
|
3026700 |
Tecal 425S |
230V |
50/60 |
700 |
21 |
|
3026100 |
Tecal 425S |
120V |
50/60 |
700 |
21 |
|
3026800 |
Tecal 425H |
230V |
50/60 |
700 |
21 |
|
3026200 |
Tecal 425H |
120V |
50/60 |
700 |
21 |
Inserts for the Tecal 425
|
Catalog Number |
Description |
Weight |
|
7032533 |
Insert 5 x ¼" probes |
13oz |
|
7032534 |
Insert 3/8" + 5/16" +1/4" + 3/16" + 1/8" probe |
13oz |
|
7032535 |
Insert 2 x 3/8" + 2 x ¼" probes |