$2,710.00

THERMOMETER, TWEENER, 110V 50/60HZ

SKU: 1502A-156 Manufacturer:Fluke
  • Description
  • Additional information

Description

Product Description

Model Name Description
1502A Tweener PRT Readout
1504 Tweener Thermistor Readout
9934-S LogWare, Single Channel, Single User
9934-M LogWare, Single Channel, Multi User

Accessories common to all models:

Accessory Description
9301 Carrying Case, fits Tweener and 12 in probe
2506 IEEE Option
2505 Spare Connector
2508 Serial Cable Kit
9320A External Battery Pack (powers 115 V ac or 12 V dc products only)
SC1-THRO Silver Care Plan, 1-yr Extended Warranty, Readouts
SC2-THRO Silver Care Plan, 2-yr Extended Warranty, Readouts
SC4-THRO Silver Care Plan, 4-yr Extended Warranty, Readouts

Specifications

1502A 1504
Temperature Range† −200 °C to 962 °C (−328 °F to 1764 °F) Any thermistor range
Resistance Range 0 Ω to 400 Ω, auto-ranging 0 Ω to 1 MΩ, auto-ranging
Probe Nominal RTPW: 10 Ω to 100 Ω
RTD, PRT, or SPRT
Thermistors
Characterizations ITS-90 subranges 4, 6, 7, 8, 9, 10, and 11
IPTS-68: R0, a, d, a4, and c4
Callendar-Van Dusen: R0, a, d, and b
Steinhart-Hart thermistor polynomial
Callendar-Van Dusen: R0, a, d, and b
Resistance Accuracy
(ppm of reading)
0 Ω to 20 Ω: 0.0005 Ω
20 Ω to 400 Ω: 25 ppm
0 Ω to 5 KΩ: 0.5 Ω
5 KΩ to 200 KΩ: 100 ppm
200 KΩ to 1 MΩ: 300 ppm
Temperature Accuracy† ±0.004 °C at −100 °C
±0.006 °C at 0 °C
±0.009 °C at 100 °C
±0.012 °C at 200 °C
±0.018 °C at 400 °C
±0.024 °C at 600 °C
±0.002 °C at 0 °C
±0.002 °C at 25 °C
±0.004 °C at 50 °C
±0.010 °C at 75 °C
±0.020 °C at 100 °C
(Using 10 KW thermistor sensor, a=0.04. Does not include probe uncertainty or characterization errors.)
Operating Temperature Range 16 °C to 30 °C 13 °C to 33 °C
Resistance Resolution 0 Ω to 20 Ω: 0.0001 Ω
20 Ω to 400 Ω: 0.001 Ω
0 Ω to 10 KΩ: 0.01 Ω
10 KΩ to 100 KΩ: 0.1 Ω
100 KΩ to 1 MΩ: 1 Ω
Temperature Resolution 0.001 °C 0.0001 °C
Excitation Current 0.5 and 1 mA, user selectable, 2 Hz 2 and 10 mA, automatically selected
Measurement Period 1 second
Digital Filter Exponential, 0 to 60 seconds time constant (user selectable)
Probe Connection 4-wire with shield, 5-pin DIN connector
Communications RS-232 serial standard
IEEE-488 (GPIB) optional
Display 8-digit, 7-segment, yellow-green LED; 0.5-inch-high characters
Power 115 VAC (±10 %), 50/60 Hz, 1 A, nominal
230 VAC (±10 %), 50/60 Hz, 1 A, nominal, specify
Size (HxWxD) 61 x 143 x 181mm (2.4 x 5.6 x 7.1 in)
Weight 1.0 kg (2.2 lb.)
Probes from Fluke Calibration
5615, 5627, 5626, 5628, 5622 5640-44, 5610-65
Calibration Accredited NIST-traceable calibration provided
†Temperature ranges and accuracy may be limited by the sensor you use.

Additional information

Weight 1 lbs