- One-year measurement uncertainty is ± 0.02% of reading from 10% to 100% with one Q-RPT, and from 1% to 100% with two Q-RPTs
- Low torque variable volume allows for pressure generation up to 200 MPa (30,000 psi) with minimal physical effort
- Separate fine adjustment tool for maximum, superfine control resolution
- User defined resolution and ready limits enable user to optimize performance based on DUT specifications
- High pressure isolation valve and pressure relief valve protect the low pressure reference transducer from over pressure when high pressure reference transducer is active
- AutoRange feature optimizes measurement and safety features for the specific range of the instrument being calibrated
- Simple rezeroing while vented at atmospheric pressure
- Simple, objective pressure “ready/not ready indicator with user adjustable criteria to ensure repeatable results among operators
- Not dependent on local gravity or ambient temperature
- Sets and reads any pressure value directly in any unit of measure, without moving weights
- Built-in priming system to fill system with test fluid and purge unwanted air to assure smooth operation
- Optional foot switch accessory allows hands-free data collection when running AutoTests
- Main Unit
- Liquid Fill Kit
- Power Supply with cord
- Set of Test Port Adapters
- Unit is shipped dry, to be filled by the customer
Hydraulic deadweight tester performance with digitalmeasurement convenience and simplicity.
|
Fluke Calibration E-DWT-H Specifications |
|
|---|---|
|
Power requirements |
12 V dc 1.2 A; 100V ac to 240V ac, 50/60 Hz |
|
Operating temperature/ humidity |
10 °C to 40 °C (50 °F to 104 °F)/ 0 % to 70 % |
|
Weight |
6531: 12 kg (26 lb) approximate 6532: 14 kg (30 lb) approximate |
|
Dimensions |
E-DWT footprint (W x D): 41.4 cm x 37.1 cm (16.3 in x 14.6 in) E-DWT height: 26.9 cm (10.6 in), 33.6 cm (13.2 in) to max variable volume handle height) |
|
Pressure ranges (Dependent on Q-RPT(s) |
20 MPa (3000 psi), 70 MPa (10,000 psi) 200 MPa (30,000 psi) |
|
Operating medium |
E-DWT-H is shipped dry and is compatible with Sebacate, silicon oils, propylene glycol, fully fluorinated liquids, partially fluorinated liquids, isopropyl alcohol, and distilled water. |
|
Reservoir capacity |
300 cc (18 in3) |
|
Variable volume displacement |
3 cc (0.18 in3) |
|
Filling and priming pump displacement |
3.7 cc (0.23 in3) |
|
TEST pressure connection |
DH500 female. Note: DH500 is a gland and collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc.. |
|
Maximum Priming Pump Pressure |
700 kPa (100 psi) |
|
Communication ports |
RS-232 (COM1, COM2) |
|
Pressure measurement |
|
|
Warm up time |
15 minute temperature stabilization recommended from cold power up |
|
Resolution |
Default: 0.01 % of active range. User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange, whichever is larger |
|
Precision |
± 0.018 % of reading or 0.0018 % of Q-RPT span, whichever is greater |
|
Predicted stability |
One year: ± 0.0075 % of reading Two year: ± 0.015 % of reading |
|
Measurement uncertainty |
One year: ± 0.02 % of reading or 0.002 % of Q-RPT span, whichever is greater Two year: ± 0.025 % of reading or 0.0025 % of Q-RPT span, whichever is greater |

Warranty: As per Manufacturer policy
100% Genuine Product
Easy and Secure Payment
Buyer Protection
Payment Options:
Be the first to review “Electronic Deadweight Tester, 200 Mpa (30K Psi) 1”