Fluke 87V identifies complex signal problems fast
The Fluke 87V Industrial Multimeter provides the resolution and accuracy to efficiently troubleshoot motor drives, plant automation, power distribution, and electromechanical equipment even in loud, high energy, and high altitude locations. A low pass filter supports accurate frequency measurements on VFDs and captures intermittents as fast as 250 µS.
Use the 87V to measure AC current up to 10 A directly, or use an optional clamp for high current measurements.frequency to 200 kHz, plus % duty cycle, resistance, continuity and diode test. And it includes a built-in thermometer so you can take temperature readings without having to carry a separate instrument, and more. Plus, with its bright, two-level backlit display you’ll be able to see the results better in dim light.
Ideal for electricians and technicians that need to measure currents, voltage or frequency on complex variable frequency drive signals. It includes the 87 V industrial multimeter, 5 x insulated screwdrivers, 3 x insulated pliers and the roll-up tool pouch.
What’s included with the Fluke IB875K
- Industrial True RMS Multimeter
- 3 x Insulated Slotted Screwdrivers
- 2 x Insulated Philips Screwdrivers
- Insulated Long Nose Pliers
- Insulated Diagonal Pliers
- Insulated Lineman Pliers
- Roll Up Pouch
Specifications | ||
Voltage DC | Maximum voltage | 1000 V |
Accuracy | ±(0.05% + 1) | |
Maximum resolution | 10 µV | |
Voltage AC | Maximum voltage | 1000 V |
Accuracy | ±(0.7% + 2) True RMS | |
AC bandwidth | 20 kHz with low pass filter; 3 db @ 1 kHz | |
Maximum resolution | 0.1 mV | |
Current DC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(0.2% + 2) | |
Maximum resolution | 0.01 µA | |
Current AC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(1.0% + 2) True RMS | |
Maximum resolution | 0.1 µA | |
Resistance | Maximum resistance | 50 MΩ |
Accuracy | ±(0.2% + 1) | |
Maximum resolution | 0.1 Ω | |
Capacitance | Maximum capacitance | 9,999 µF |
accuracy | ±(1% + 2) | |
Maximum resolution | 0.01 nF | |
Frequency | Maximum frequency | 200 kHz |
Accuracy | ±(0.005% + 1) | |
Maximum resolution | 0.01 Hz | |
Duty cycle | Maximum duty cycle | 99.9% |
Accuracy | ±(0.2% per khz + 0.1%) | |
Maximum resolution | 0.1% | |
Temperature measurement | –200.0 °C – 1090 °C –328.0 °F – 1994.0 °F excluding probe | |
80 BK temperature probe | –40.0 °C – 260 °C –40.0 °F – 500 °F, 2.2 °C or 2% whichever is greater | |
Conductance | Maximum conductance | 60.00 nS |
Accuracy | ±(1.0% + 10) | |
Maximum resolution | 0.01 nS | |
Diode | Range | 3 V |
Resolution | 1 mV | |
Accuracy | ±(2% + 1) | |
Duty cycle range | Accuracy | Within ±(0.2% per kHz + 0.1%) |
Environmental Specifications | ||
Operating temperature | –20 °C to + 55 °C | |
Storage temperature | –40 °C to + 60 °C | |
Humidity (without condensation) | 0% – 90% (0 °C – 35 °C) 0% – 70% (35 °C – 55 °C) | |
Operating Altitude | 2000 m | |
Safety Specifications | ||
Overvoltage category | EN 61010–1 to 1000 V CAT III, 600V CAT IV | |
Agency approvals | UL, CSA, TÜV, VDE listed | |
Mechanical and General Specifications | ||
Size | 201 x 98 x 52 mm (with holster) | |
Weight | 355 g 624 g – with holster | |
Display | Digital | 6000 counts updates 4/sec. 19,999 counts in high–resolution mode |
Analog | 32 segments, updates 40/sec | |
Frequency | 19,999 counts, updates 3/sec at > 10 Hz | |
Warranty | Lifetime | |
Battery Life | Alkaline | ~400 hours typical, without backlight |
Shock | 1 Meter drop per IEC 61010–1:2001 | |
Vibration | Per MIL–PRF–28800 for a Class 2 instrument |