Kanomax particle sensors with analog output are designed to fit into your existing monitoring system, or they can be used as a stand-alone unit to monitor a critical area when connected to an alarm or controller.
Measurement Specifications | |
Measuring Particle Size | 3718-A: 0.3µm and 0.5µm; 3719-A: 0.5µm and 5.0µm |
Counting Efficiency | 50% +/- 20% at smallest particle channel |
Coincidental Loss | Less than 10% at 500,000 particles/cf |
Zero Count | Less than 1 count / 5 minutes |
General Specifications | |
Flow Rate | 1.0 cfm (28.3 l/min) |
Light Source | Laser diode |
Interfaces | RS485 (Modbus capability); Analog output (4-20mA) |
Enclosure | Stainless steel |
Power Supply | DC24V (over network hub) or 9 to 28V (AC adapter) |
Output | Particle counts, error status, alarm output |
Vacuum Source | External pump |
Operating Environment | 10 to 30°C (50 to 86°F)/20 to 85%RH with no condensation |
Dimensions | W 5.2 x D 3.7 x H 3.1 inch (133 x 95 x 80 mm) |
Weight | 2.2 lbs (1.0 kg) |
Standards | ISO21501-4, JIS B9921, CE-mark |
Warranty | 3 year |
Standard Accessories | User’s Manual – English (1),User’s Manual – Japanese (1), AC adapter (1), Zero filter (1), Adapter 1 (1),Adapter 2 (1), Tygon tube 1 (1), Setting software DVD – 3718/19-40 (1) |
Be the first to review “Particle Sensor (1CFM, 0.3/0.5 μm, RS485 Modbus, 4-20mA output)”