Techne Secondary Standard Probe |
| |
Our Secondary standard probes offer proven and reliable performance for your critical temperature measurements. The SSP660 can be used with our Tecal Accu-Temp indicator or any comparable unit and the WSTC1450 with a higher accuracy 6 1/2 or 8 1/2 digit multimeter. The SSP660 includes a hard carrying case and ITS-90 calibration (NIST traceable) certificate with coefficients. The SSP660 gives a higher level of accuracy and stability than our WSP probes making it a good selection for use as a secondary standard reference thermometer. The WSTC1450 is assembled into a high purity alumina sheath for long term stability and accuracy. Calibration is supplied via a CD with an Excel spreadsheet containing polynomials for mV to temperature conversion. |
Choose one for your application |
|
|
Probe Size | 1/4″ OD x 18″L with handle and 72″ of PTFE coated 22ga. 4 wire lead | 1/4″OD x 18″ long alumina sheath with attached 1/4″OD x 8″ long diameter reference junction with 72″ inches of lead |
Calibrated Temperature Range | -196 to 660°C | 0 to 1450°C |
Description | SSP660 Probe w/ carrying case. Includes ITS90 calibration with | WSTC1450 reference type S thermocouple with half junction reference and calibration supplied on CD containing polynomials for mV to temperature conversion |
Resistance at 0°C | 25.5Ω | N/A |
Temperature coefficient | 0.003925 | N/A |
Stability | RO Typical drift <0.01°C after | Long term; typical drift <0.5°C at 1100°C after 250 hours |
Repeatability | RO typical drift < 0.005°C after cycling from -196 to 670°C |
|
Calibrated uncertainty K=2 | –196°C ±0.050°C | 500 to 1450°C ±1.80°C |
ORDERING INFORMATION: | Catalog number: 7002795 | Catalog number: 7032802 |
Be the first to review “Techne Secondary Standard Probe”