- Measurements include conductivity, resistivity, salinity, total dissolved solids and temperature.
- With YSI’s exclusive accurate Resistance Ratio TechnologyTM (RRT), the 3200 continually compares the resistance of the sample to a calibrated resistor in the circuit.
- Ensure the most accurate measurements with NIST-traceable conductivity cells with built-in temperature sensors and exceed USP24 specifications.
- To ensure even greater accuracy, utilize the YSI conductivity calibration solutions.
Features:
- Highly accurate Resistance Ratio Technology
- Multipoint calibration
- High and low alarms
- Linear and nonlinear temperature compensation options
- Data storage and logging
YSI 3200 Instrument General Specifications
Alarm | Yes | |
Calibration | Multipoint, up to 5 points | |
Cell Connector | 7-pin Mini Din | |
Cell constant, cm-1 | 0.001 – 100 cm-1 | |
Cell Information Storage | 6 Configurations | |
Certifications | UL, CSA, CE | |
Clock | Yes | |
Data Storage | 100 points | |
Dimensions | Size | 9 by 9.5 by 4.4 in (22.9 by 24.1 by 11.2 cm) |
Weight | 2.6 lbs (1.1 kg) | |
Display | Graphic LCD | |
Operating Environment | 95% relative humidity, non-condensing | |
Outputs | RS-232 | |
Platinizing Capability | Included | |
Power | AC, 115V, 220V | |
Temperature Compensation | Method | Linear, nonlinear |
Reference temperature °C | 0 – 100 | |
Temperature. Coefficient | 0 – 10%/°C, nonlinear | |
Water Resistance | The Model 5000 and 5100 are designed exclusively for indoor use and are NOT waterproof. | |
Warranty | 2-year instrument |
YSI 3200 Specifications (Instrument Only)
Conductance | |||
Range | Accuracy | Resolution | Frequency |
0 – 0.9999 µS | ± 0.30% full scale | 0.0001 µS | 40 Hz |
0.950 – 9.999 µS | ± 0.20% full scale | 0.001 µS | 80 Hz |
9.50 µS – 99.99 µS | ± 0.10% full scale | 0.01 µS | 290 Hz |
95.0 – 999.9 µS | ± 0.10% full scale | 0.1 µS | 1010 Hz |
950 – 9999 µS | ± 0.10% full scale | 1 µS | 1010 Hz |
9.50 – 99.99 mS | ± 0.10% full scale | 0.01 mS | 1010 Hz |
95.0 – 999.9 mS | ± 0.30% full scale | 0.1 mS | 1460 Hz |
0.95 – 3.00 S | ± 1.0% full scale | 0.01 S | 1460 Hz |
Resistance | |||
Range | Accuracy | Resolution | Frequency |
0 – 9.999 Ω | ± 0.2% full scale | 0.001Ω | 1460 Hz |
0 – 99.99 Ω | ± 0.1% full scale | 0.01 Ω | 1460 Hz |
0 – 999.9 Ω | ± 0.1% full scale | 0.1 Ω | 1010 Hz |
0 – 9.999 kΩ | ± 0.1% full scale | 0.001 kΩ | 1010 Hz |
0 – 99.99 kΩ | ± 0.1% full scale | 0.01 kΩ | 1010 Hz |
100.0 – 999.9 kΩ | ± 0.2% full scale | 0.1 kΩ | 290 Hz |
1.00 – 9.99 MΩ | ± 0.5% full scale | 0.01 MΩ | 80 Hz |
10.0 – 29.9 MΩ | ± 1% full scale | 0.1 MΩ | 40 Hz |
Range | Accuracy | Resolution | |
Salinity | 0-80 ppt (NaCl) | ±.1 ppt | 0.1 ppt |
TDS | 0-19,999 mg/L | 0.50% | 1 mg/L |
Temperature | -5 – 100°C | ±0.1°C | 0.01°C |
Conductivity Cell General Measurement Ranges
Conductivity Cell General Specifications
Be the first to review “YSI 3200-230 Conductivity Meter; 230 VAC”