The MIC-5010 is a portable, rugged, easy-to-use, 5 kV class insulation test meter. The MIC-5010 has a wide measurement range, selectable test voltages, and excellent safety features.
Insulation resistance measurement features:
- User-selectable test voltage any from 50 to 1,000 V (with 10 V resolution) and 1,000 to 5,000 V (with 25 V resolution)
- Continuous indication of measured insulation resistance or leakage current
- Audio beeps every five-second periods to indicate measurement progress
- Indication of actual test voltage during measurement
- 1.2 mA and 3 mA test current
- Step voltage insulation resistance measurement (SV)
- Selectable test times (T1, T2 , T3 from 1 to 600 seconds) to measure one or two absorption coefficients
- Adjustable measuring time to 99′ 59’’
- Polarization index (PI) and dielectric absorption ratio (DAR) measurements
- Dielectric Discharge calculation (DD)
- High-noise immunity; digital filters function for measurements in high noise enviroment (10 s, 30 s , 60 s)
- Continuity measurement of protective connections and equipotential bonding in accordance with EN 61557-4
- Adjustable limits for measured resistance RISO and RCONT.
- Measurement of capacitance during the measurement of RISO
- DC and AC voltage measurement in the range of 0 to 600 V
- Meter conforms to EN 61557
Safety First Features.
- CAT-IV 600V, CAT-III 1000V
- Automatic discharge of tested object after the insulation resistance measurement completes to remove high-voltage charge remaining due to cable capacitance
- Protection against measuring live objects, measurement will not proceed
- Measurements with test leads up to 66 ft / 20 m length
Construction
- Backlit keyboard and display
- Rugged IP 54 case – Protected from water spray from any direction
Power Supply
- Power supply from battery packs, low battery warning indicator, built-in fast charger
Memory Features
- 990 cells of memory (11,880 records) to save test results
- Data transmission to a PC via USB cable
Insulation resistance measurement (two-lead)
Measurement range acc. to IEC 61557-2: 50k to 15.0 TΩ (IISOnom = 1.2 mA or 3 mA)
Range | Resolution | Accuracy |
0.0 to 999 kΩ | 1 kΩ | ±(3% m.v. + 10 digits) |
1.00 to 9.99 MΩ | 0.01 MΩ | |
10.0 to 99.9 MΩ | 0.1 MΩ | |
100 to 999 MΩ | 1 MΩ | |
1.00 to 9.99 GΩ | 0.01 GΩ | |
10.0 to 99.9 GΩ | 0.1 GΩ | |
100 to 999 GΩ | 1 GΩ | ±(3.5% m.v. + 10 digits) |
1.00 to 9.99 TΩ | 0.01 TΩ | ±(7.5% m.v. + 10 digits) |
10.0 to 15.0 TΩ | 0.1 TΩ | ±(10% m.v. + 10 digits) |
Values of measured resistance depending on measurement voltage
Voltage UISO | Measurement range |
250 V | 500 GΩ |
500 V | 1.00 TΩ |
1,000 V | 2.00 TΩ |
2,500V | 5.00 TΩ |
5,000V | 15.0 TΩ |
Measurement of leakage current
Range | Resolution | Accuracy |
0 to ILmax | m, µ, n [A] | Calculated basing on resistance measurements |
· ILmax – maximum current at short circuit of leads,
· resolution and units result from the measurement range of individual insulation resistance.
Step voltage insulation resistance measurement
Target voltage | Measurement voltage sequence |
1 kV | 200, 400, 600, 800, 1,000 V |
2.5 kV | 0.5, 1, 1.5, 2, 2.5 kV |
5 kV | 1, 2, 3, 4, 5 kV |
· duration of each “step” adjustable from 30s to 5mins
· measurement result for each voltage step is stored in memory
Continuity measurement of protective connections and equipotential bonding with 200 mA current
Measurement range acc. to EN 61557-4: 0.12 to 999 Ω
Range | Resolution | Accuracy |
0.00 to 19.99 Ω | 0.01 Ω | ±(2% m.v. + 3 digits) |
120.0 to 199.9 Ω | 0.1 Ω | |
200 to 999 Ω | 1 Ω | ±(4% m.v. + 3 digits) |
· Voltage on open terminals: 4 to 24 V
· Output current at R < 2 Ω: Imin > 200 mA (ISC: 200 to 250 mA)
· Compensation of test lead resistance
· Current flowing in both directions, mean value of resistance is displayed
Capacity measurement
Range | Resolution | Accuracy |
1 to 999 nF | 1 nF | ±(5% m.v. + 5 digits) |
1.00 to 49.99 µF | 0.01 µF |
· Capacity measurement result is displayed after the RISO measurement
DC and AC voltage measurement
Range | Resolution | Accuracy |
0.0 to 29.9 V | 0.1 V | ±(2% m.v. + 20 digits) |
30.0 to 299.9 V | 0.1 V | ±(2% m.v. + 6 digits) |
300 to 600 V | 1 V | ±(2% m.v. + 2 digits) |
The acronym “m.v.” stands for a “measured reference value”.